Back to Home
Femtosecond beam path for a nonlinear laser-scanning microscope
Technical guides

Nonlinear microscopy

How to build a two-photon microscope

Excitation confined to the focus, so no pinhole is needed and scattered light still counts as signal. Everything then follows from one scaling law.

Signal ∝ P²/(τ·f)

Absorption is quadratic in intensity. Shorter pulses and lower repetition rates buy signal at the same average power — and the same heating.

No pinhole

Only the focal volume is excited, so every collected photon is in-focus by definition. Collect non-descanned and keep the scattered light.

Dispersion is the tax

The objective and scan optics stretch the pulse. A 100 fs pulse arriving as 300 fs costs two thirds of your signal.

Interactive excitation-budget lab

Where your two-photon signal actually goes

Pulse at the sample

149 fs

Peak power

1.6 kW

Pulse energy

0.25 nJ

Relative signal

0.67×

Dispersion penalty

100 fs at the laser → 149 fs at the focus · 67% of the signal retained

Balanced
Transform-limited100%
With your GDD67%

Duty cycle, focal confinement and dose are in a workable range for this configuration.

Focal spot ≈ 299 nm radial · 1080 nm axialSignal ∝ P²/(τ·f) — relative to 920 nm, 100 fs, 80 MHz, 20 mW, NA 1.0

Why there is no pinhole

Excitation per z-plane, normalised to the focal plane · NA 1.00 · 920 nm · zR389 nm

Widefield

Excited
Entire double cone
Detected
Everything, in-focus and out
Out-of-focus bleaching

Full-depth bleaching

Confocal

Excited
Entire double cone
Detected
Pinhole rejects out-of-focus
Out-of-focus bleaching

Full-depth bleaching — the pinhole discards the signal, it does not prevent the damage

Two-photon

Excited
Focal volume only
Detected
All of it, non-descanned
Out-of-focus bleaching

Confined to the focal plane

-4-202410Defocus z (µm)
One-photon — widefield and confocalTwo-photon
53%of excitation delivered through a 10 µm stack lands inside the focal plane (two-photon)
8%of excitation delivered through a 10 µm stack lands inside the focal plane (one-photon)

The one-photon line is flat because total power is conserved as the beam spreads: every plane above and below the focus absorbs the same amount. That is the photobleaching a confocal pinhole cannot prevent — it only discards the emission afterwards.

A first-order model for comparing configurations, not a substitute for measuring pulse width at the focus. Real systems add higher-order phase, objective transmission that falls steeply past 1000 nm, and wavelength-dependent scattering in tissue.

Femtosecond sources

The beam path

Six stages between the laser and the detector

Source

Femtosecond pulses, 80 MHz. Rep rate sets pulse energy at a given average power.

Femtosecond lasers

Power control

An AOM sets power per plane and blanks the beam during flyback. Add an isolator to protect the laser.

AOM & shutters

Beam expansion

Fill the objective back aperture. Underfilling lowers the effective NA and throws away resolution and signal.

Beam expanders

Scan & relay

Scan lens and tube lens conjugate the mirror to the pupil, so the beam pivots without walking off it.

Cage & relay optics

Z and sample

Piezo objective or sample positioning for stacks. Closed-loop if plane spacing must be metrology.

Piezo stages

Collection

Dichroic and short-pass blocking, collected non-descanned. Scattered fluorescence still carries signal.

Filters & dichroics

Detectors for two-photon are PMTs or hybrid detectors placed close to the objective. We do not supply those — for descanned or widefield channels alongside the nonlinear path, see the scientific cameras.

Sources

80 MHz femtosecond lasers from the catalog

Wavelength follows the fluorophore; average power has to survive the scan path before it reaches the sample.

Blue/green indicators

FS-H-780A

780 nm

GFP, CFP and DAPI-class fluorophores sit near their two-photon peak around 780 nm.

Output / average power
1-50 mW
Repetition rate
80±2 MHz
Single pulse energy (nJ)
0.1-0.6
View product
Deep tissue, red indicators

FS-H- 1030B

1030 nm

Longer wavelengths scatter less. Headroom for power lost to the scan path and the objective.

Output / average power
500-3000 mW
Repetition rate
80±2 MHz
Single pulse energy (nJ)
6.25-30
View product
Maximum penetration

FS-H- 10 64B

1064 nm

The most average power of the three, for thick specimens and multi-point excitation.

Output / average power
500-5000 mW
Repetition rate
80±2 MHz
Single pulse energy (nJ)
6.25-62.5
View product

Before you order

Six things that decide whether it works

Specify the system

Tell us the fluorophore and the depth.

We will size the wavelength, power and dispersion budget with you before anything is quoted.