Back to Home
CARS energy-level diagram beside a spectrum showing the anti-Stokes signal blue-shifted away from the fluorescence background

Complete system guide

Building a CARS microscope

Coherent anti-Stokes Raman scattering turns a molecule’s own bonds into contrast. Design the two colours, the delay, the filters, and the detector as one instrument — and know what the background is doing.

17 min read Three-scheme design lab Label-free imaging

Two colours make the contrast

The frequency difference between pump and Stokes selects the chemical bond. Nothing is added to the sample.

Coherence makes it fast

Driven vibrations radiate in phase, so a CARS image can be acquired orders of magnitude faster than a spontaneous Raman map.

The signal lands where fluorescence cannot

A blue-shifted anti-Stokes photon is separable by filtering — but not from the non-resonant four-wave-mixing background.

The decision in one sentence

Pick the vibration first; the pulse width follows from its linewidth.

Almost every hard choice in a CARS system — laser architecture, synchronization, delay hardware, detector — is downstream of two numbers: the Raman shift you need and how wide that band is. A 2 ps pulse and a 100 fs pulse differ by a factor of twenty in bandwidth, and only one of them matches a molecular vibration.

01

Name the bond

CH₂ at 2 845 cm⁻¹, C≡D at 2 100 cm⁻¹, amide I near 1 655 cm⁻¹, or an unknown fingerprint region.

02

Bound the spectrum

One fixed band, a few hundred wavenumbers of hyperspectral range, or a whole fingerprint per pixel.

03

Set the dose

Power at the sample, repetition rate, dwell time, and what the specimen survives for the length of a scan.

1 · How the signal forms

Drive the vibration, then scatter off it.

01

Two colours, one difference

The pump and Stokes beams are focused together. Only when ω_p − ω_S matches a vibrational frequency do the molecules in the focus start oscillating in phase.

02

A driven, coherent ensemble

Unlike spontaneous Raman, the vibrations are not independent. They are phase-locked to the driving field, which is where the signal gain comes from.

03

A third field scatters

The probe field — usually the pump beam again — scatters off that coherence and generates a new field at 2ω_p − ω_S.

04

Signal builds coherently

Because the emitters radiate in phase, intensity scales with the square of the number of oscillators and emerges as a directional, laser-like beam.

05

Blue of the pump

The anti-Stokes photon has more energy than any of the input photons, so it lands where fluorescence cannot reach.

06

Non-resonant background rides along

Electrons respond to the same fields instantly and produce four-wave mixing whether or not a vibration is resonant. That term never goes away.

The anti-Stokes frequency

ωas = 2ωp − ωS
ω_p − ω_S
the vibrational frequency being driven — this is the Raman shift
ω_as
always higher in energy than any input photon, so it lands blue of the pump
I_CARS
scales as the pump intensity squared times the Stokes intensity
χ⁽³⁾
a resonant term plus a real, non-resonant term that never switches off

Because the detected quantity is |χ⁽³⁾|², the resonant and non-resonant terms interfere in amplitude. That interference — not noise — is what makes CARS peaks look shifted and asymmetric compared with a spontaneous Raman spectrum of the same molecule.

A CARS spectrum is not a Raman spectrum.

Peak positions can appear shifted by several wavenumbers, weak bands can invert into dips, and concentration dependence is not linear at low concentration because the resonant term interferes with the background. If the result has to be compared with a Raman library, plan the phase-retrieval step at design time — it constrains whether you need a spectrum at every pixel.

2 · Excitation schemes

Three ways to make the difference frequency, three different instruments.

The scheme decides how you change the Raman shift — by tuning a laser, by moving a stage, or not at all because every shift is recorded at once.

Narrowband picosecond

Sources
Two synchronized ps beams — typically one laser plus an OPO
Resolution
≈ 5–15 cm⁻¹, set by the pulse bandwidths
Best for
Fast single-band imaging: lipids at 2 845 cm⁻¹, deuterium labels, protein amide

Tradeoff: The cleanest spectral selectivity and the best signal-to-background per milliwatt, but changing the Raman shift means tuning a laser.

Spectral focusing

Sources
One femtosecond source, both arms chirped by matched glass or gratings
Resolution
≈ 7–25 cm⁻¹, set by the chirped duration and how well the two chirps match
Best for
Hyperspectral imaging over a few hundred wavenumbers without retuning

Tradeoff: A delay stage replaces wavelength tuning, so the stage step, repeatability and chirp matching become the spectral axis.

Multiplex / broadband

Sources
Narrowband pump plus a broadband Stokes continuum
Resolution
Whatever the spectrograph and detector deliver
Best for
A full fingerprint spectrum per pixel where chemical identity is unknown

Tradeoff: Every wavenumber is recorded at once, but photons are spread across the array and each pixel needs far longer dwell.

Complete path

Every stage between the laser and the number you record.

01

Source

Shared clock, both colours

02

Chirp / filter

Set the bandwidth

03

Delay

Temporal overlap

04

Combine

Dichroic, collinear

05

Focus + scan

High-NA objective

06

Reject

Block pump and Stokes

07

Detect

PMT or spectrograph

The hardest component to source is the second colour. A CARS microscope needs two synchronized wavelengths whose difference lands on the vibration. In practice that is an optical parametric oscillator pumped by the same laser, a soliton-shifted fiber branch, or a photonic-crystal-fiber continuum. Whichever route you take, it is a system-level decision made before any single component is ordered — tell us the target Raman shift and we can work backwards from it.

3 · Pulses + delay

Bandwidth you cannot use is dose you cannot spend.

A femtosecond pulse delivers enormous peak power, but a molecular vibration only accepts the fraction of that bandwidth sitting inside its linewidth. The rest still reaches the sample, still generates non-resonant background, and still contributes to photodamage.

01

Pulse width

A 2 ps Gaussian pulse has about 7 cm⁻¹ of bandwidth; a 100 fs pulse has about 147 cm⁻¹. Condensed-phase Raman bands are 10–20 cm⁻¹ wide, so most of a femtosecond pulse is spent outside the band it is exciting.

02

Temporal overlap

Pump and Stokes must arrive within a fraction of a pulse width. Below that, the CARS signal simply vanishes — which is also the most reliable way to find time zero.

03

Synchronization jitter

Two separately mode-locked lasers need active timing stabilisation. An OPO pumped by the same laser, or two arms of one source, avoids the problem entirely.

04

Repetition rate

At 80 MHz the pulse energy is low and the average power dominates heating; at 1 MHz the pulse energy is 80× higher and peak-intensity damage arrives first. Both routes reach the same sample.

05

Dispersion

Every lens, dichroic, and the objective itself stretch the pulse. What matters is the pulse width at the focus, not at the laser output.

06

Chirp matching

In spectral focusing the two arms must carry identical chirp. If they do not, the difference frequency sweeps across the pulse and resolution degrades whatever the delay stage does.

Transform limit

Δν̃ ≈ 14.7 / τps cm⁻¹

For a Gaussian pulse: 2 ps gives about 7 cm⁻¹, 500 fs about 29 cm⁻¹, 100 fs about 147 cm⁻¹. Compare against a condensed-phase Raman linewidth of 10–20 cm⁻¹ and the picosecond preference for narrowband CARS follows directly.

Delay as a spectral axis

Δt = 2Δx / c

With a retroreflector, 1 µm of stage travel is 6.67 fs of delay. In spectral focusing that becomes a wavenumber step: a 100 fs pulse chirped to 2 ps tunes at roughly 74 cm⁻¹ per picosecond, so one wavenumber is about 2 µm of travel.

In spectral focusing, stage specifications become spectroscopy specifications.

Minimum incremental motion sets the finest wavenumber step. Repeatability sets whether a spectrum recorded on the way out matches the one recorded on the way back. Backlash appears as a shift between scan directions, and settling time is dead time in every pixel. None of that matters in a fixed-band microscope — and all of it matters the moment the delay stage is your wavelength knob.

Important: the pulse width that matters is the one at the focus. A high-NA objective plus dichroics and scan optics can stretch a sub-100 fs pulse substantially, which changes both peak intensity and effective resolution. Measure it at the sample plane, or budget for the pre-compensation that lets you.

4 · Detection chain

Decide what a pixel means before choosing the detector.

One number per pixel at megahertz rates, or a whole spectrum per pixel at kilohertz rates. That single choice determines the filters, the scan strategy, the acquisition electronics, and how long a field of view takes.

Rejection first

The anti-Stokes signal can be many orders of magnitude weaker than the excitation. Filter selection follows from the computed anti-Stokes wavelength, not from a catalog default.

Collection geometry

Forward detection needs a condenser matched to the excitation NA; epi detection reuses the objective and sees a different subset of scatterers.

Bandwidth vs. spectrum

A PMT gives microsecond dwell and one channel. An array gives hundreds of channels and millisecond dwell. There is no configuration that gives both.

Sampling the focus

Pixel size should sample the diffraction-limited spot, which in CARS is set by the excitation wavelengths — not by the shorter anti-Stokes wavelength you detect.

Forward CARS (F-CARS)

Condenser + filters + PMT or photodiode

Why it helps

Phase matching is naturally satisfied over the focal volume, giving the strongest signal from thin, transparent specimens

What to verify

Needs optical access on both sides of the sample. Signal is dominated by objects larger than roughly the excitation wavelength.

Epi CARS (E-CARS)

Same objective, dichroic, PMT

Why it helps

Works on thick or opaque samples and suppresses the bulk solvent contribution

What to verify

Backward signal comes mainly from small scatterers and interfaces, plus back-reflected forward signal — the contrast mechanism is not the same.

Point detector · PMT / APD

Single-channel, photon-counting or analogue

Why it helps

Megahertz bandwidth for pixel dwell times in the microsecond range — the only practical choice for video-rate raster scanning

What to verify

One number per pixel. Anything spectral has to come from tuning, not from the detector.

Spectrograph + array

Cooled line CCD or back-illuminated sCMOS

Why it helps

A whole anti-Stokes spectrum per position, which is what makes multiplex CARS and background retrieval possible

What to verify

Read noise, dark current, well depth and readout rate now set the pixel dwell time — a full spectrum per pixel costs milliseconds, not microseconds.

5 · Design lab

Put the Raman shift, the pulse width, and the detector on one budget.

Set the vibration you want and the lab returns the Stokes wavelength you have to source, where the signal lands, whether silicon can see it, what the excitation resolves, and what the focus is doing to your sample. Switch schemes to see how the same target changes the hardware.

Model the experiment

CARS design lab

Two narrowband picosecond beams parked on the 2 845 cm⁻¹ CH₂ stretch — the classic single-band CARS microscope.

Where the three colours land

silicon cutoff ≈ 1 000 nmanti-Stokes650 nmpump797 nmStokes1,031 nm400700100013001600
Anti-Stokes signal650 nm
Stokes beam required1,031 nm
Spectral resolution10.4 cm⁻¹
Diffraction-limited spot0.81 µm
Pump peak intensity28.5 GW/cm²
Relative CARS yield
The excitation resolves the band you named

Excitation resolution is 10.4 cm⁻¹ against a 15 cm⁻¹ target. Shorter pulses buy peak power and lose spectral selectivity — condensed-phase bands are typically 10–20 cm⁻¹ wide, so a sub-picosecond pulse spends most of its bandwidth outside the band.

Silicon detection with a clean filter gap

The signal at 650 nm sits 147 nm blue of the pump. A silicon detector covers it, and the blue shift is what keeps one-photon fluorescence — which is always red of the excitation — out of the detection band.

Dose is in the range live samples usually tolerate

40 mW combined at the sample, 0.31 nJ per pump pulse, 28.5 GW/cm² at focus. Damage thresholds are sample-specific: verify by scanning power and repetition rate independently and watching for a signal that no longer scales as expected.

Pump pulse energy0.31 nJ
Stokes pulse energy0.19 nJ
Pump–anti-Stokes gap147 nm

Gaussian pulses and a diffraction-limited Airy focus are assumed throughout. Resolution in spectral-focusing mode is the best case for perfectly matched chirp on both arms. Relative yield uses the resonant CARS scaling P̄p²·P̄S / (f²·τ²·A³), normalised to the CH₂ preset — it compares configurations and is not an absolute photon count. It also ignores the non-resonant background, which scales differently and often decides whether a weak band is usable at all.

6 · Background + artifacts

The background is a physical signal, not noise.

Averaging does not reduce it, and no filter separates it. Every practical CARS system deals with the non-resonant term either optically, at the point of excitation, or numerically, after the spectrum is recorded.

Non-resonant background

What it looks like

A featureless four-wave-mixing pedestal from the electronic response of solvent, glass and immersion medium — present even with no vibration in range.

Control

Picosecond excitation, off-resonance reference images, polarization CARS, time-delayed probing, or a spectral fit that models the interference explicitly.

Dispersive lineshapes

What it looks like

Peaks look shifted, asymmetric or dip below the baseline because the resonant and non-resonant terms interfere in amplitude, not intensity.

Control

Retrieve the imaginary part of χ⁽³⁾ — maximum-entropy or Kramers–Kronig — before comparing anything with a spontaneous Raman library.

Photodamage

What it looks like

Signal drops during a scan, or morphology changes between frames; often nonlinear in power and worse at low repetition rate.

Control

Scan average power and repetition rate independently, move to a fresh field, and confirm the signal still scales as expected against pump and Stokes power.

Imperfect overlap

What it looks like

Signal is weak but the spectrum looks right — or the signal appears and disappears with focus.

Control

Optimise spatial overlap at the sample plane and temporal overlap on delay; both beams must be collinear through the full pupil, not just at the objective back aperture.

Pump and Stokes leakage

What it looks like

Excitation light reaching the detector inherits every fluctuation of the laser.

Control

Short-pass and bandpass filtering chosen from the actual anti-Stokes wavelength, plus spatial and angular rejection. Verify with each beam blocked in turn.

Two-photon fluorescence

What it looks like

A broad background that survives when the beams are temporally separated.

Control

Block one beam, change the delay far beyond overlap, and compare — any surviving signal is not CARS.

Block the Stokes

Everything that remains is not CARS.

Move off resonance

A few hundred wavenumbers away isolates the non-resonant pedestal.

Separate in time

Delay the beams past overlap; the coherent signal disappears, backgrounds may not.

Scale the power

Resonant CARS follows pump² × Stokes. Anything that does not is something else.

7 · Relevant products

Build the instrument stack from the vibration outward.

Catalog-backed starting points for ultrafast excitation, the delay arm that carries the spectral axis, beam routing, spectral dispersion and detection. They are components, not a turnkey microscope — the second colour, the objective, the scanner and the filter set still need to be designed around your target band.

Ultrafast excitation

Spectral-focusing platform1030 nm

FS-H-1030B

An 80 MHz femtosecond source at 1030 nm with 0.5–3 W of average power — the usual starting point when one laser has to feed both arms of a chirped-pulse CARS setup.

Output / average power
500-3000 mW
Repetition rate
80±2 MHz

Qualify: Confirm pulse width and spectral bandwidth at the sample after the objective, available power for a second colour, and the trigger or clock output the detection chain will use.

View product
1064 nm fundamental1064 nm

FS-H-10 64B

An 80 MHz femtosecond source at 1064 nm with 0.5–5 W, for systems that build the Stokes branch on the fundamental and derive the pump from parametric conversion.

Output / average power
500-5000 mW
Repetition rate
80±2 MHz

Qualify: Check pulse width, power stability over an imaging session, and whether the beam quality supports diffraction-limited focusing at high NA.

View product
Narrowband picosecond branch1030 nm

FL-H-1030-PS

A picosecond 1030 nm fiber platform at 1–20 W, for the narrowband scheme where spectral selectivity — not peak power — is the design goal.

Output / average power
1~20 W
Beam / notes
Picosecond Pulsed Infrared Laser

Qualify: Ask for the actual pulse width and optical bandwidth, not just the picosecond classification: 2 ps and 20 ps are different instruments for this application.

View product

Delay line and beam gating

Zolix LAK20-60 motorized linear stage
Spectral-focusing delay armZolix

LAK20-60

In spectral focusing the delay stage is the spectral axis. A submicron-class compact stage covers the full chirped bandwidth in a couple of millimetres of travel while keeping each step small compared with a wavenumber.

Mechanical travel
20 mm
Double-pass delay window
≈ 133 ps
Minimum incremental motion
≤ 1 µm · ≈ 6.7 fs
Fine-subdivision resolution
0.1 µm · ≈ 0.67 fs
Repositioning accuracy
≤ ±1 µm
Maximum speed
20 mm/s

Qualify: The stage is the moving platform, not the delay line. Add a retroreflector or roof-mirror assembly, a compatible controller, and confirm settling time, approach direction and real optical repeatability in the assembled geometry. If you also need a long-window pump–probe arm, the 150 mm LA150-60 covers roughly a nanosecond.

View Zolix stage
CNI AOM-I acousto-optic modulator
Beam gating and modulationCNI

AOM-I acousto-optic modulator

Fast gating keeps light off the sample between scan lines, sets a modulation pattern for lock-in detection, or reduces an 80 MHz train to a rate the sample survives.

  • 450–700 nm
  • 100 MHz centre frequency
  • Up to 4 MHz pulse repetition
  • 80% diffraction efficiency
  • TTL / analogue control

Wavelength check: The catalog AOM-I is specified for 450–700 nm and typical optical power below 1 W with a 1.0 mm beam — a visible-band device. For a near-infrared CARS excitation beam, confirm wavelength coverage, single-pulse energy, dispersion, extinction ratio and rise time with the supplier before assuming it fits.

Zolix beam-routing optomechanics

Pump and Stokes have to stay overlapped in space and angle through the full objective pupil, over hours. Mount stability is not a detail here — a drift of a fraction of a beam diameter is a drift in signal.

Zolix NMC25.4 beam combining and steering
Beam combining and steering

NMC25.4

Centre kinematic mount for Ø25–25.4 mm optics — the dichroic and the fold mirrors that bring pump and Stokes onto a common axis.

  • ±3° angular range
  • Ø23 mm clear aperture
  • 25 mm optical-axis height
View product
Zolix NMUM25.4 spatial-overlap control
Spatial-overlap control

NMUM25.4

Five-axis mount for a 25.4 mm optic, for the collinearity adjustment that has to hold both beams overlapped through a high-NA objective.

  • ±2.5 mm X/Y
  • 6 mm Z travel
  • ±5° tip/tilt
View product
Zolix NPH50 repeatable beam height
Repeatable beam height

NPH50

Compact base-mounted holder for Ø12 mm posts, keeping the delay arm and the combining optics at one table height.

  • Ø12 mm post
  • 50 mm overall height
  • Base-mounted holder
View product

Spectrographs for hyperspectral CARS

Multiplex + multi-track detection

HiperS-320i

Toroidal on-axis mirrors keep the focal plane free of astigmatism, which matters when a broadband anti-Stokes spectrum must stay sharp across a whole line array or when signal and reference occupy separate tracks.

  • 320 mm focal length
  • f/4.4 aperture
  • Aberration-corrected focal plane
  • Triple grating turret

Qualify: Choose grating, blaze, slit width, input f/number and detector format together — the anti-Stokes span from the design lab tells you how much of the focal plane you actually need.

View instrument
General hyperspectral CARS

Omni-λ300i

A practical 320 mm Czerny–Turner platform when moderate resolution, throughput and grating flexibility have to coexist on one budget.

  • 320 mm focal length
  • f/4.2 aperture
  • Triple grating turret
  • Multiple entrance/exit ports

Qualify: Work out the wavenumber-per-pixel on the actual sensor before buying resolution. In CARS the limit is usually photons per pixel, not grating dispersion.

View instrument

Detectors for the anti-Stokes signal

sCCD01AM scientific detector
Multiplex CARS spectra

sCCD01AM

A 2048 × 1 deep-cooled scientific CCD with 15 µm pixels, a 30.7 mm active length, 459 ke⁻ full well and 95% quantum efficiency at 800 nm — a line detector built for a spectrograph focal plane rather than an imaging port.

Sensor
E2V CCD261
Resolution
2048 × 1
Pixel Size
15 µm × 15 µm
Frame Rate
12 fps @ 2048 × 1
Interface
USB3

Qualify: The 300–1000 nm range covers anti-Stokes signals from all the usual pump/Stokes pairs. Check cooling (55 °C below ambient), the 22 e⁻ read noise against your per-pixel photon budget, and whether 12 fps is compatible with your scan pattern.

View camera
sMAX04BM-CL100 scientific detector
Hyperspectral + widefield

sMAX04BM-CL100

A 2048 × 2048 back-illuminated sCMOS with Camera Link and 108 fps full-frame readout, for imaging CARS geometries and for spectra that need more than one detector track.

Sensor
GSENSE2020BSI (sCMOS)
Resolution
4.2 MP (2048×2048)
Pixel Size
6.5 µm × 6.5 µm
Frame Rate
108 fps @ 2048×2048; 108 fps @ 1024×1024
Interface
Camera Link

Qualify: Verify ROI timing, external trigger behaviour and read noise at the gain you will actually use. Full-frame rate is not the same as usable spectra per second once dwell and scan overhead are included.

View camera

Want the whole CARS chain reviewed?

Bring the target Raman shift and band linewidth, the sample and its damage limits, the field of view and frame rate you need, and whether the result has to be comparable with a spontaneous Raman spectrum.

8 · Specification brief

Specify a measurement, not a shopping list.

These are the inputs that expose the interfaces between laser, mechanics, optics, detector and software before anything is quoted.

01

Target vibration

Raman shift in cm⁻¹, band linewidth, expected cross-section relative to a known reference such as CH₂ at 2 845 cm⁻¹.

02

Excitation scheme

Narrowband, spectral focusing, or multiplex — and therefore how the Raman shift is changed between measurements.

03

Sources

Pump and Stokes wavelengths, pulse widths at the sample, repetition rate, synchronization method and residual timing jitter.

04

Delay

Overlap range, step size in wavenumbers, repeatability, and whether the delay stage is scanned during acquisition or only set.

05

Sample

Thickness, refractive index, immersion medium, damage threshold, motion, and whether epi access is required.

06

Filtering

Anti-Stokes wavelength, required blocking of pump and Stokes, and the optical density needed at the detector.

07

Detection

Point detector or spectrograph plus array; bandwidth, dwell time, quantum efficiency at the anti-Stokes wavelength, and read noise.

08

Acquisition

Pixel dwell, field size, frames averaged, background reference strategy, and how spectra will be phase-retrieved.

9 · Common questions

Short answers before the design review.

What is CARS microscopy?+

Coherent anti-Stokes Raman scattering microscopy is a label-free imaging technique. Two synchronized laser beams — pump and Stokes — are focused into a sample; when their frequency difference matches a molecular vibration, the vibration is driven coherently and a third field scatters off it to produce a directional signal at a new, blue-shifted wavelength. Contrast comes from the molecule's own chemical bonds, so no dye or fluorescent label is needed.

How is CARS different from spontaneous Raman?+

Spontaneous Raman scattering is incoherent and extremely weak, which is why a Raman map can take hours. In CARS the vibrations are driven in phase, so the emitters add coherently and the signal can be several orders of magnitude stronger — enough for video-rate imaging. The price is a non-resonant background, interference-distorted lineshapes, and two synchronized pulsed sources instead of one CW laser.

Why does the signal come out at a shorter wavelength?+

The anti-Stokes photon carries the energy of two pump photons minus one Stokes photon, so it is higher in energy than any input photon. Because one-photon fluorescence is always red-shifted from the excitation, a blue-shifted signal can be separated from fluorescence with a filter — the main practical advantage over Stokes-side techniques.

What is the non-resonant background and can it be removed?+

It is four-wave mixing from the instantaneous electronic response of the sample and its surroundings, present whether or not a vibrational resonance is being driven. It cannot be filtered away, because it appears at exactly the anti-Stokes wavelength. It can be suppressed — picosecond excitation, polarization CARS, epi detection, time delay — or accounted for by retrieving the imaginary part of the third-order susceptibility from a measured spectrum.

Do I need picosecond or femtosecond pulses?+

Picosecond pulses match the width of a typical condensed-phase Raman band and give the best ratio of resonant signal to non-resonant background. Femtosecond pulses give more peak power and more bandwidth than a single band can use, but with matched chirp on both arms — spectral focusing — they recover picosecond-like resolution while letting a delay stage tune the Raman shift.

Should I detect with a PMT or with a spectrograph and camera?+

Use a point detector when the Raman shift is fixed and the goal is a fast image: microsecond dwell times and megahertz bandwidth are only available that way. Use a spectrograph and an array detector when you need the spectrum itself — for unknown chemistry, for overlapping bands, or because you intend to phase-retrieve the lineshape. Many systems carry both.

How does CARS compare with SRS?+

Stimulated Raman scattering measures the intensity change of the excitation beams themselves rather than a new colour. It has no non-resonant background and its spectra match spontaneous Raman directly, but it requires modulation and lock-in detection because the signal is a small fractional change on a bright beam. CARS needs no lock-in chain and produces a background-free-looking wavelength, at the cost of the interference. The laser and scanning hardware are largely shared.

Research basis

This guide is an engineering overview. Representative primary work behind the scheme comparison, pulse-width argument, spectral focusing, and background retrieval:

  1. 1. Zumbusch, Holtom & Xie, “Three-dimensional vibrational imaging by coherent anti-Stokes Raman scattering,” Physical Review Letters 82, 4142 (1999).
  2. 2. Cheng, Volkmer, Book & Xie, “An epi-detected coherent anti-Stokes Raman scattering (E-CARS) microscope with high spectral resolution and high sensitivity,” Journal of Physical Chemistry B 105, 1277 (2001).
  3. 3. Hellerer, Enejder & Zumbusch, “Spectral focusing: high spectral resolution spectroscopy with broad-bandwidth laser pulses,” Applied Physics Letters 85, 25 (2004).
  4. 4. Vartiainen, Rinia, Müller & Bonn, “Direct extraction of Raman line-shapes from congested CARS spectra,” Optics Express 14, 3622 (2006).
  5. 5. Evans & Xie, “Coherent anti-Stokes Raman scattering microscopy: chemical imaging for biology and medicine,” Annual Review of Analytical Chemistry 1, 883 (2008).

Continue designing

Go deeper on the spectral chain or the timing.