Gated, framing & streak cameras
TIMART ST10, T40 & T40-HDR Streak Cameras
- Photocathode options: S20: 200–850 nm; BB: 200–900 nm; S25: 350–900 nm
- ST10 synchronised sweep: 40–300 MHz
- Triggered sweep windows: 1 ns–1 ms, model dependent
- Slit adjustment: 0–3 mm, manual

TIMART cameras were developed by the Xi’an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences. They are offered here as a separately attributed camera family for spectroscopy integration.
Best suited to
Time-resolved photoluminescence, scintillator emission and fast discharge measurements.
Selecting a configuration
Specify single-shot or repetitive operation first, then the time window and spectral-image length. Confirm tube, photocathode and readout configuration together; a tube figure is not a guaranteed system response.
Measurement & configuration
- ST10 lens-coupled or fibre-faceplate readout.
- Single- or dual-MCP ST10 configurations.
- Spectrograph integration for wavelength–time images.
Plan the signal path
- 1Optical slit
- 2Streak-tube sweep
- 3Time-resolved image
Specifications
| Photocathode options | S20: 200–850 nm; BB: 200–900 nm; S25: 350–900 nm |
|---|---|
| ST10 synchronised sweep | 40–300 MHz |
| Triggered sweep windows | 1 ns–1 ms, model dependent |
| Slit adjustment | 0–3 mm, manual |
Published family comparison
| Family | Typical tube resolution | Cathode slit | Triggered repetition |
|---|---|---|---|
| ST10 | ≤2 ps | 8 mm | Single shot or <10 kHz |
| T40 | 50 ps | 35 mm | Single shot or <1 kHz |
| T40-HDR | 10 ps | 30 mm | Single shot or <1 kHz |
Configuration note: Ordering models include ST10-1L/-2L, ST10-1F/-2F, T40-1F and T40-1F-HDR. Use model-specific figures; the source overview and detailed table differ. The malformed synchronous sweep-range unit is omitted.
Related gated, framing & streak cameras
IsCMOS Intensified Scientific Cameras
Time-gated plasma emission, LIBS and weak transient luminescence.
ExploreICCD Intensified High-Resolution Cameras
Narrow spectral features, transient emission and high spatial resolution gated measurements.
ExploreFC Ultrafast Framing Cameras
Non-repeatable combustion, discharge, detonation and plasma events.
ExploreConfigure TIMART ST10 / T40 / T40-HDR
Send the wavelength interval, signal level, timing and sample requirements for your experiment.