Back to Home
Goniometer25-theta

Opto-Mechanics · Cryogenic Piezo Stages - Goniometers

Goniometer25-theta

Cryogenic theta goniometer; 25 x 25 mm footprint; 6.6 deg tilt range; 41 mm pivot distance above plate; 200 g payload; resistive closed-loop feedback around 0.2-0.5 mdeg resolution.

View quote list
Datasheet (PDF) ↗CAD Model (STEP) ↗

Specifications

Active axesRx
Footprint × hight25 x 25 x 12.5 mm
Mass28 g
Work environmentDefault:1.4 K~ 400 K, 1e-7 mbar, 35 Tesla
MainbodyDefault: Pure Ti
WiresPhosphor Bronze Twisted Paired Wires, 20cm
Pin materialsPolyster (glass fiber filled), BeCu
Pins numberDrive -2 pins, Sensor- 3 pins
Fine Tune Resolution @2 K*0.5 u°
Step Size (min) @300 K*50 u°
Travel range~6.6 °
Max. Velocity @300 K~1°/s
Max. Load200 g
Dynamic Force2.2 N
Rotation center to top plate41 mm
SensorResistive Sensor
Sensor range6.6 °
Sensor resolution0.2 m°

Toepassingen

Kantelcorrectie na afkoeling

Herstel de vlakligging van het monster nadat differentiële krimp de houder heeft gekanteld.

Tweeassige tip/tilt

Stapel theta- en phi-eenheden om rond een gemeenschappelijk virtueel draaipunt boven de bovenplaat te sturen.

Invalshoek in krappe geometrie

Stel de bundelinval in binnen een cryostaat waar geen ruimte is om in plaats daarvan de optiek te bewegen.

Specifying this part

What this model has to survive, and what it needs around it

Fixed pivot above the top plate

Each unit tilts about a centre that sits a fixed height above its own mounting surface — 41 to 66 mm depending on model. Stacking a theta and a phi unit gives dual-axis tip/tilt about one shared virtual pivot, which is why they are specified in matched pairs.

Heat load at the cold plate

Stick-slip inertial drives dissipate energy per step and draw no holding current once stopped, so the steady-state load on the mixing chamber is the wiring loom rather than the motor. Anchor the loom at each temperature stage and thermalise the moving body with an FTC copper braid.

Non-magnetic construction

Titanium bodies and BeCu fasteners throughout, rated for operation to 35 T and down to 30 mK with the .ULT option, at 2 × 10⁻¹¹ mbar with .UHV. A single steel screw substituted at installation is enough to compromise a high-field assembly.

Full series comparison, controller pairing and mounting hardware on the cryogenic piezo stages overview, or work through the selection with the configurator.