Back to Home

Pulse energy, repetition rate, pulse width

EO-266-G266 nm deep-UV Q-switched pulsed laser

266 nm

EO-266-G is a 266 nm Pulsed laser. Used for scientific research, biology, instrumentation, laser marking. Key specifications: output / average power 100~500 mW, pulse energy 4~100 uJ, repetition rate 1 Hz-25 kHz. View specifications, datasheet and enquiry options from Precisometer.

Specifications

EO-266-G

15 specification rows

Output / average power100~500 mW
Pulse energy4~100 uJ
Repetition rate1 Hz-25 kHz
Pulse width<9 ns @1 kHz
Power stability (rms, 4 hours, ±3℃)<5%, <3%
Pointing stability over temperature (urad/℃)<20
Beam divergence, full angle (mrad)<3
Beam diameter (mm)~1
Warm-up time (minutes)<15
Cooled methodAir cooled or water cooled
Operating temperature (℃)15-30
Power supply (24VDC)PSU-EO-G (24V/13A)
Wavelength (nm)266±1
Operating modeActively Q-switched
Beam height from base plate (mm)80, 56

Application context

Why this design is used

Applications named on this model’s datasheet, expanded with the model’s optical specifications.

Scientific research

EO-266-G: 266 nm emission determines compatible optics, coatings and detectors for Scientific research. The catalogued beam parameter (<3) helps predict how the source will focus and propagate through the instrument.

Biology

EO-266-G: 266 nm emission determines compatible optics, coatings and detectors for Biology. The catalogued beam parameter (<3) helps predict how the source will focus and propagate through the instrument.

Instrumentation

EO-266-G: 266 nm emission determines compatible optics, coatings and detectors for Instrumentation. The catalogued beam parameter (<3) helps predict how the source will focus and propagate through the instrument.

Laser marking

EO-266-G: Laser marking depends on how optical energy is delivered to the target; 266 nm emission governs absorption, while the listed temporal parameter (4~100 uJ) governs peak intensity and heat deposition.

System integration

Integration and compatibility

Use these checks to connect EO-266-G to the surrounding optical, mechanical and control system. Every value shown is taken from this model's own specification record.

Thermal management
Cooled method: Air cooled or water cooled · Operating temperature (℃): 15-30
Provide the stated cooling method and keep the laser-head base thermally coupled; temperature drift moves wavelength, output power and beam pointing.
Beam parameters at the output
Beam divergence, full angle (mrad): <3 · Beam diameter (mm): ~1
Size the downstream optics from the stated beam diameter and divergence rather than from the aperture of the housing, and confirm the polarisation state before specifying isolators or polarising optics.

Technical FAQ

Integration questions for EO-266-G

What cooling does the EO-266-G require?

Cooled method: Air cooled or water cooled · Operating temperature (℃): 15-30. Provide the stated cooling method and keep the laser-head base thermally coupled; temperature drift moves wavelength, output power and beam pointing.

What beam does the EO-266-G deliver?

Beam divergence, full angle (mrad): <3 · Beam diameter (mm): ~1. Size the downstream optics from the stated beam diameter and divergence rather than from the aperture of the housing, and confirm the polarisation state before specifying isolators or polarising optics.

Need this configured?

Share wavelength, operating mode, output power or pulse energy, linewidth or pulse width, delivery, control, and safety requirements and we will qualify the matching configuration and lead time.

Open laser selector

Model selection

Nearby series comparison

Compare EO-266-G with the closest available models in the Pulsed Lasers family. Confirm option-dependent values on the final configuration before ordering.

Selection parameterEO-266-G (current)MPL-Q-266MPL-Q1-266
Wavelength266 nm266 nm266 nm
Optical power / pulse energy100~500 mW1~30 mW10 mW
Operating mode / pulse width1 Hz-25 kHz3-12 kHz1 kHz
Beam quality<3Near TEM00Near TEM00