Back to Home

Spectral cameras & detectors

DField-VS-A1/A2/A3 Cooled Spectral CCDs

Scientific CCD readout heads for weak-signal spectroscopy across 200–1100 nm. Select A1 for UV-sensitive acquisition, A3 for near-infrared sensitivity or the deeply cooled A2 for low-light integration.

DField-VS-A1/A2/A3 Cooled Spectral CCDs
DField-VS-A1 / A2 / A3

Best suited to

Spectrograph readout for Raman, fluorescence and emission measurements.

Selecting a configuration

Match active width and pixel pitch to the spectrograph focal plane. Compare cooling at the stated ambient temperature and distinguish a camera sensor array from a complete fibre spectrometer.

Measurement & configuration

  • 16-bit acquisition over USB 2.0.
  • Spectral binning for signal collection.
  • A1/A3 trigger input and shutter-control output.

Plan the signal path

  1. 1Dispersed spectrum
  2. 2Cooled CCD array
  3. 3Binned spectrum

Specifications

DField-VS-A1 / A2 / A3 specifications
Wavelength interval200–1100 nm
A1/A3 integration10 ms–30 min
A1/A3 dynamic range9000
A1/A3 supply12 V DC
A1/A3 size / weight99 × 119 × 78 mm / 1 kg

Published sensor configurations

Published sensor configurations
ModelPixelsPixel pitchCooling at 25 °C ambient
DField-VS-A12048 × 12212 µm−10 °C
DField-VS-A22000 × 25615 µm−60 °C
DField-VS-A32048 × 6414 µm+5 °C

Configuration note: The A1/A3 headline cooling statement is generic; this comparison follows the model-specific table. DField-VS-A4 remains available on its existing product page.

Configure DField-VS-A1 / A2 / A3

Send the wavelength interval, signal level, timing and sample requirements for your experiment.

Request a configuration