Opto-Mechanics · Cryogenic Piezo Stages - Closed-loop Ultra scanners
Scanner25-x.Ultra.ULT
Closed-loop Ultra X-axis scanner; 25 mm cryogenic scanner profile; integrated high-linearity capacitive displacement sensors with sub-nanometer resolution class; ultra-low-temperature ULT grade down to 30 mK; use with MC-ArchimedesLT.03.Ultra controller.
Datasheet (PDF) ↗Specifications
Full specifications for this product are available on request. Contact us for the datasheet.
Applications
Quantitative imaging at base temperature
Capacitive feedback ties image pixels to real distances rather than to applied voltage.
Cryogenic AFM and STM raster
Deliver linear, drift-corrected scan under a probe head operating below 4 K.
Repeat-visit measurement
Return to the same coordinate across thermal cycles and long measurement campaigns.
Specifying this part
What this model has to survive, and what it needs around it
What the capacitive loop buys you
An integrated capacitive displacement sensor closes the loop at 0.5 nm and lets the controller compensate thermal drift, piezoelectric creep and self-heating actively. Worth the extra wiring only if the measurement holds position for hours; if the sample is re-found optically each run, open-loop costs less heat.
Reading the HV / UHV / ULT suffixes
The suffix sets the environmental preparation, not the mechanics: HV, UHV, ULT and UHV.ULT variants differ in materials, cabling and bake compatibility. Match it to the chamber the stage will live in — the mechanical specification is identical across them.
Heat load at the cold plate
Stick-slip inertial drives dissipate energy per step and draw no holding current once stopped, so the steady-state load on the mixing chamber is the wiring loom rather than the motor. Anchor the loom at each temperature stage and thermalise the moving body with an FTC copper braid.
Non-magnetic construction
Titanium bodies and BeCu fasteners throughout, rated for operation to 35 T and down to 30 mK with the .ULT option, at 2 × 10⁻¹¹ mbar with .UHV. A single steel screw substituted at installation is enough to compromise a high-field assembly.
Full series comparison, controller pairing and mounting hardware on the cryogenic piezo stages overview, or work through the selection with the configurator.
Related products
Scanner25-x.Ultra
Closed-loop Ultra X-axis scanner; 25 mm cryogenic scanner profile; integrated high-linearity capacitive displacement sensors with sub-nanometer resolution class; standard cryogenic environment grade; use with MC-ArchimedesLT.03.Ultra controller.
Scanner25-x.Ultra.HV
Closed-loop Ultra X-axis scanner; 25 mm cryogenic scanner profile; integrated high-linearity capacitive displacement sensors with sub-nanometer resolution class; high-vacuum HV environment grade; use with MC-ArchimedesLT.03.Ultra controller.
Scanner25-x.Ultra.UHV
Closed-loop Ultra X-axis scanner; 25 mm cryogenic scanner profile; integrated high-linearity capacitive displacement sensors with sub-nanometer resolution class; ultra-high-vacuum UHV environment grade; use with MC-ArchimedesLT.03.Ultra controller.
Scanner25-x.Ultra.UHV.ULT
Closed-loop Ultra X-axis scanner; 25 mm cryogenic scanner profile; integrated high-linearity capacitive displacement sensors with sub-nanometer resolution class; combined UHV plus ultra-low-temperature grade down to 30 mK; use with MC-ArchimedesLT.03.Ultra controller.