Opto-Mechanics · Ambient Piezo Stages - Objective scanners
Carrier.OB400.C
Objective Z scanner; 400 um Z travel; 5 nm closed-loop resolution; 165 Hz resonant frequency at 150 g load; M25 x 0.75 objective mount.
Datasheet (PDF) ↗Specifications
| Active axes | Z |
|---|---|
| Mass | 350 g |
| Travel range | 400 μm |
| Pitch/Yaw | ±30 µrad |
| Linearity | 0.0003 |
| Resolution(open loop) | 0.8 nm |
| Resolution(closed-loop) | 5 nm |
| Repeatability | 10 nm |
| Sensor Type | Capacitive Sensor |
| Response Time | 24 ms |
| Stiffness | 0.25 N/um |
| Resonant Frequency(no load) | 380 Hz |
| Resonant Frequency(150 g) | 165 Hz |
| Controller | MC-Archimedes.N series |
Applications
Fast Z-sectioning
Move the objective instead of the sample, so optical sections can be stepped without disturbing live specimens.
Autofocus and focus tracking
Hold focus against thermal drift over acquisitions lasting minutes to hours.
Light-sheet and multiphoton stacks
Acquire Z-stacks at a rate set by the scanner rather than by the microscope frame.
Specifying this part
What this model has to survive, and what it needs around it
Driving the objective rather than the sample
Moves the lens along the optical axis for focus stacks and z-series, leaving a liquid sample or mounted cell undisturbed. Travel options run 100, 200 and 400 µm; settling time rather than range is usually the limiting specification.
Vacuum and non-magnetic versions
Most of the range is orderable as .HV (high vacuum), .UHV (ultra-high vacuum) or .NM (non-magnetic). The suffix changes materials, cabling and bake compatibility rather than the mechanics, so specify it against the chamber the stage will live in.
Full series comparison, controller pairing and mounting hardware on the ambient piezo stages overview, or work through the selection with the configurator.
Related products
Carrier.OB.C
Objective scanner base with M25 x 0.75 objective thread; supports microscope autofocus integration and objective piezo positioning modules.
Carrier.OB100.C
Objective Z scanner; 100 um Z travel; 1 nm closed-loop resolution; 250 Hz resonant frequency at 150 g load; M25 x 0.75 objective mount.
Carrier.OB200.C
Objective Z scanner; 200 um Z travel; 3 nm closed-loop resolution; 220 Hz resonant frequency at 150 g load; M25 x 0.75 objective mount.
Carrier.OBHL100.C
High-load objective Z scanner; 100 um Z travel; 1 nm closed-loop resolution; 560 Hz resonant frequency at 150 g load for high-bandwidth focusing.