Back to Home
Carrier.OB400.C

Opto-Mechanics · Ambient Piezo Stages - Objective scanners

Carrier.OB400.C

Objective Z scanner; 400 um Z travel; 5 nm closed-loop resolution; 165 Hz resonant frequency at 150 g load; M25 x 0.75 objective mount.

View quote list
Datasheet (PDF) ↗

Catalog context

Where Carrier.OB400.C fits

Carrier.OB400.C is listed in the Ambient Piezo Stages - Objective scanners family. The local catalog record provides these first selection fields: Active axes: Z; Mass: 350 g; Travel range: 400 μm.

Specifications

Active axesZ
Mass350 g
Travel range400 μm
Pitch/Yaw±30 µrad
Linearity0.0003
Resolution(open loop)0.8 nm
Resolution(closed-loop)5 nm
Repeatability10 nm
Sensor TypeCapacitive Sensor
Response Time24 ms
Stiffness0.25 N/um
Resonant Frequency(no load)380 Hz
Resonant Frequency(150 g)165 Hz
ControllerMC-Archimedes.N series

Application context

Why this design is used

Fast Z-sectioning

Move the objective instead of the sample, so optical sections can be stepped without disturbing live specimens.

Autofocus and focus tracking

Hold focus against thermal drift over acquisitions lasting minutes to hours.

Light-sheet and multiphoton stacks

Acquire Z-stacks at a rate set by the scanner rather than by the microscope frame.

System integration

Integration and compatibility

Use these checks to connect Carrier.OB400.C to the surrounding optical, mechanical and control system. Every value shown is taken from this model's own specification record.

Travel and resolution
Travel range: 400 μm · Resolution(open loop): 0.8 nm
Separate travel from resolution and both from repeatability. Repeatability is what an automated sequence actually depends on, and it is the figure most often absent from a headline specification.
Drive and feedback
Resolution(closed-loop): 5 nm · Controller: MC-Archimedes.N series
Pair open-loop mechanics with the specified high-voltage or inertial driver, and closed-loop variants with the controller their capacitive, strain-gauge or encoder feedback requires. Matching connectors do not guarantee electrical compatibility.
Load and mounting
Resonant Frequency(no load): 380 Hz
Check the load rating in the orientation you will use it — a vertical or cantilevered load is a moment, not a mass — and match the base to an M6 pattern on a 25 mm grid or a 1/4"-20 pattern on a 1 inch grid.

Technical FAQ

Integration questions for Carrier.OB400.C

What travel and resolution does the Carrier.OB400.C provide?

Travel range: 400 μm · Resolution(open loop): 0.8 nm. Separate travel from resolution and both from repeatability. Repeatability is what an automated sequence actually depends on, and it is the figure most often absent from a headline specification.

Which controller does the Carrier.OB400.C require?

Resolution(closed-loop): 5 nm · Controller: MC-Archimedes.N series. Pair open-loop mechanics with the specified high-voltage or inertial driver, and closed-loop variants with the controller their capacitive, strain-gauge or encoder feedback requires. Matching connectors do not guarantee electrical compatibility.

What load can the Carrier.OB400.C carry, and how does it mount?

Resonant Frequency(no load): 380 Hz. Check the load rating in the orientation you will use it — a vertical or cantilevered load is a moment, not a mass — and match the base to an M6 pattern on a 25 mm grid or a 1/4"-20 pattern on a 1 inch grid.

Specifying this part

What this model has to survive, and what it needs around it

Driving the objective rather than the sample

Moves the lens along the optical axis for focus stacks and z-series, leaving a liquid sample or mounted cell undisturbed. Travel options run 100, 200 and 400 µm; settling time rather than range is usually the limiting specification.

Vacuum and non-magnetic versions

Most of the range is orderable as .HV (high vacuum), .UHV (ultra-high vacuum) or .NM (non-magnetic). The suffix changes materials, cabling and bake compatibility rather than the mechanics, so specify it against the chamber the stage will live in.

Full series comparison, controller pairing and mounting hardware on the ambient piezo stages overview, or work through the selection with the configurator.

Model selection

Nearby series comparison

Compare Carrier.OB400.C with the closest available models in the Ambient Piezo Stages - Objective scanners family.

Selection parameterCarrier.OB400.C (current)Carrier.OB200.XY.CCarrier.OBHL100.C
Travel / adjustment400 μm200 X 200 μm100 μm
Resolution / sensitivity0.8 nm2 nm0.3 nm