Back to Home
S500.X.C

Opto-Mechanics · Ambient Piezo Stages - Standard scanner modules

S500.X.C

Standard flexure-guided X-axis piezo scanner; 500 um X-axis travel; capacitive closed-loop feedback; closed-loop resolution 1.5 nm class.

View quote list
Datasheet (PDF) ↗

Catalog context

Where S500.X.C fits

S500.X.C is listed in the Ambient Piezo Stages - Standard scanner modules family. The local catalog record provides these first selection fields: Active axes: X; Mass: 240 g; Travel range(open-loop): 600 um.

Specifications

Active axesX
Mass240 g
Travel range(open-loop)600 um
Travel range(closed-loop)500 um
Resolution(open-loop)0.7 nm
Resolution(closed-loop)1.5 nm
Linearity0.0003
Pitch/Yaw±6/±6 urad
Resonant Frequency200 Hz
Max. Pull Force10 N
Max. Payload (horizontal mounting)500 g
Sensor TypeCapacitive Sensor
Max Driving Voltage150 V
ControllerMC-Archimedes.N series

Application context

Why this design is used

Closed-loop raster scanning

Capacitive feedback linearises the scan, removing the hysteresis that distorts open-loop images.

Scanning probe microscopy

Provide the fine XY or XYZ scan under an AFM, STM or near-field head.

Confocal and Raman mapping

Step a sample through a focus in reproducible increments so map pixels correspond to real coordinates.

System integration

Integration and compatibility

Use these checks to connect S500.X.C to the surrounding optical, mechanical and control system. Every value shown is taken from this model's own specification record.

Travel and resolution
Travel range(open-loop): 600 um · Travel range(closed-loop): 500 um
Separate travel from resolution and both from repeatability. Repeatability is what an automated sequence actually depends on, and it is the figure most often absent from a headline specification.
Drive and feedback
Resolution(open-loop): 0.7 nm · Resolution(closed-loop): 1.5 nm
Pair open-loop mechanics with the specified high-voltage or inertial driver, and closed-loop variants with the controller their capacitive, strain-gauge or encoder feedback requires. Matching connectors do not guarantee electrical compatibility.
Load and mounting
Max. Payload (horizontal mounting): 500 g
Check the load rating in the orientation you will use it — a vertical or cantilevered load is a moment, not a mass — and match the base to an M6 pattern on a 25 mm grid or a 1/4"-20 pattern on a 1 inch grid.

Technical FAQ

Integration questions for S500.X.C

What travel and resolution does the S500.X.C provide?

Travel range(open-loop): 600 um · Travel range(closed-loop): 500 um. Separate travel from resolution and both from repeatability. Repeatability is what an automated sequence actually depends on, and it is the figure most often absent from a headline specification.

Which controller does the S500.X.C require?

Resolution(open-loop): 0.7 nm · Resolution(closed-loop): 1.5 nm. Pair open-loop mechanics with the specified high-voltage or inertial driver, and closed-loop variants with the controller their capacitive, strain-gauge or encoder feedback requires. Matching connectors do not guarantee electrical compatibility.

What load can the S500.X.C carry, and how does it mount?

Max. Payload (horizontal mounting): 500 g. Check the load rating in the orientation you will use it — a vertical or cantilevered load is a moment, not a mass — and match the base to an M6 pattern on a 25 mm grid or a 1/4"-20 pattern on a 1 inch grid.

Specifying this part

What this model has to survive, and what it needs around it

Closed-loop travel is shorter than open-loop travel

The loop reserves headroom to correct within, so the usable range drops when the sensor is engaged: an S100.XY.C covers 120 × 120 µm open-loop against 100 × 100 µm closed-loop, and an S200 covers 240 × 240 against 200 × 200. Specify against the closed-loop number.

Flexure guiding, capacitive sensing

A piezo stack acting through flexure hinges with a capacitive sensor reading directly against the moving surface — no sliding contact, no backlash, and millisecond-class response.

Vacuum and non-magnetic versions

Most of the range is orderable as .HV (high vacuum), .UHV (ultra-high vacuum) or .NM (non-magnetic). The suffix changes materials, cabling and bake compatibility rather than the mechanics, so specify it against the chamber the stage will live in.

Full series comparison, controller pairing and mounting hardware on the ambient piezo stages overview, or work through the selection with the configurator.

Model selection

Nearby series comparison

Compare S500.X.C with the closest available models in the Ambient Piezo Stages - Standard scanner modules family.

Selection parameterS500.X.C (current)S300.Z.C
Travel / adjustment600 um350 um
Resolution / sensitivity0.7 nm0.5 nm
Mounting interface500 g500 g