S100.XYZ.C

Opto-Mechanics · Ambient Piezo Stages - Standard scanner modules

S100.XYZ.C

Standard flexure-guided XYZ-axis piezo scanner; 100 x 100 x 100 um XYZ travel; capacitive closed-loop feedback; closed-loop resolution 0.5 nm Z / 1 nm XY.

View quote list
Datasheet (PDF) ↗

Specifications

Active axesX Y Z
Mass450 g
Travel range(open-loop)120 × 120 × 120 μm
Travel range(closed-loop)100 × 100 × 100 μm
Resolution(open-loop)0.2 nm
Resolution(closed-loop)1 nm
Linearity0.0003
Resonant Frequency(X/Y/Z, no load)400/ 350/ 300 Hz
Stiffness0.5 N/um
Max. Payload (horizontal mounting)500 g
Sensor TypeCapacitive Sensor
Max Driving Voltage150 V
CableD-Sub 3W3
ControllerMC-Archimedes.N series

Toepassingen

Closed-loop rasterscannen

Capacitieve terugkoppeling lineariseert de scan en verwijdert de hysterese die open-loop beelden vervormt.

Scanning-probemicroscopie

Levert de fijne XY- of XYZ-scan onder een AFM-, STM- of nabijveldkop.

Confocale en Raman-mapping

Stapt een monster reproduceerbaar door de focus, zodat mappixels met echte coördinaten overeenkomen.

Specifying this part

What this model has to survive, and what it needs around it

Closed-loop travel is shorter than open-loop travel

The loop reserves headroom to correct within, so the usable range drops when the sensor is engaged: an S100.XY.C covers 120 × 120 µm open-loop against 100 × 100 µm closed-loop, and an S200 covers 240 × 240 against 200 × 200. Specify against the closed-loop number.

Flexure guiding, capacitive sensing

A piezo stack acting through flexure hinges with a capacitive sensor reading directly against the moving surface — no sliding contact, no backlash, and millisecond-class response.

Vacuum and non-magnetic versions

Most of the range is orderable as .HV (high vacuum), .UHV (ultra-high vacuum) or .NM (non-magnetic). The suffix changes materials, cabling and bake compatibility rather than the mechanics, so specify it against the chamber the stage will live in.

Full series comparison, controller pairing and mounting hardware on the ambient piezo stages overview, or work through the selection with the configurator.