FTIR spectrometers: interferograms and spectral resolution
Match optical path range, source, beam splitter and detector to the required infrared measurement.
- 01Broadband source
- 02Interferometer
- 03Interferogram
- 04Fourier spectrum
How it works
- An FTIR instrument records an interferogram as optical path difference changes. A Fourier transform converts that signal into a spectrum.
- Maximum optical path difference is related to spectral resolving capability. Display point spacing or zero filling is not an independent improvement in physical resolution.
- Source, beam splitter, windows and detector jointly determine the usable spectral band. A broad detector response alone does not establish full instrument coverage.
- Transmission, reflection and microscope measurements require different sample interfaces and reference procedures.
What to compare
| Parameter | Check in the specification |
|---|---|
| Resolution | Keep path range, apodisation and reporting definition with the value. |
| Working band | Check every optical element and the selected detector configuration. |
| Signal quality | Compare acquisition time, background, number of scans and sample conditions. |
| Sample interface | Confirm accessories, apertures and reference geometry. |
Integration
- Choose the sample method before selecting or comparing accessory packages.
- Measure background and sample under compatible environmental and alignment conditions.
- Do not infer a particular detector material or internal interferometer layout from the FTIR label alone.