Back to Home

Technology guide

FTIR spectrometers: interferograms and spectral resolution

Match optical path range, source, beam splitter and detector to the required infrared measurement.

  1. 01Broadband source
  2. 02Interferometer
  3. 03Interferogram
  4. 04Fourier spectrum
Conceptual operating path.

How it works

  • An FTIR instrument records an interferogram as optical path difference changes. A Fourier transform converts that signal into a spectrum.
  • Maximum optical path difference is related to spectral resolving capability. Display point spacing or zero filling is not an independent improvement in physical resolution.
  • Source, beam splitter, windows and detector jointly determine the usable spectral band. A broad detector response alone does not establish full instrument coverage.
  • Transmission, reflection and microscope measurements require different sample interfaces and reference procedures.

What to compare

ParameterCheck in the specification
ResolutionKeep path range, apodisation and reporting definition with the value.
Working bandCheck every optical element and the selected detector configuration.
Signal qualityCompare acquisition time, background, number of scans and sample conditions.
Sample interfaceConfirm accessories, apertures and reference geometry.

Integration

  • Choose the sample method before selecting or comparing accessory packages.
  • Measure background and sample under compatible environmental and alignment conditions.
  • Do not infer a particular detector material or internal interferometer layout from the FTIR label alone.

Related products

References

Discuss the configuration. Send the sample format, spectral range, resolution and detection requirements.