Omni-S Imaging Spectrometers & Monochromators

Three imaging-corrected grating platforms with 320, 500 and 750 mm focal lengths. Interchangeable triple-grating turrets support wavelength scanning and array-camera spectroscopy; the 500S adds motorised focus adjustment for each grating.

Omni-S Imaging Spectrometers & Monochromators
Omni-λ300s / Omni-500S / Omni-λ750S

Best suited to

Research spectroscopy requiring a choice between throughput, resolution and camera imaging.

Selecting a configuration

Start with the required wavelength interval and detector pixel size. Specify the grating, slit width and entrance/exit geometry together: a longer focal length narrows the band captured on a fixed detector.

Measurement & configuration

  • Triple-grating turret; gratings selected separately.
  • Manual slit supplied; motorised options depend on port.
  • USB control and optional configurations for UV or infrared optics.

Plan the signal path

  1. 1Entrance slit
  2. 2Grating dispersion
  3. 3Slit or array readout

Specifications

Omni-λ300s / Omni-500S / Omni-λ750S specifications
Focal plane30 × 14 mm
Grating size68 × 68 mm
Slit width0.01–3 mm
Slit heights2, 4 or 14 mm
Wavelength step, 1200 lines/mm0.005 nm
Wavelength accuracy, 1200 lines/mm±0.2 nm
Repeatability, 1200 lines/mm±0.025 nm
CommunicationUSB 2.0; RS-232 configuration dependent

Platform comparison at 1200 lines/mm

Platform comparison at 1200 lines/mm
PlatformFocal length / apertureSlit resolution¹CCD resolution²Weight
Omni-λ300s320 mm / f/4.20.08 nm0.173 nm20 kg
Omni-500S500 mm / f/6.50.05 nm0.11 nm27 kg
Omni-λ750S750 mm / f/9.70.03 nm0.054 nm40 kg

¹ 10 µm slit near 435.83 nm. ² Published values for 15 µm CCD pixels. Spectral coverage depends on grating and detector; mechanical scan limits do not imply optical response.

Configure Omni-λ300s / Omni-500S / Omni-λ750S

Send the wavelength interval, signal level, timing and sample requirements for your experiment.