Timing & signal acquisition

DCS900PC Time-Correlated Photon Counting

A time-correlated single-photon counting acquisition unit covers fluorescence decay, phosphorescence and afterglow measurements with selectable time bins and a wide acquisition window.

DCS900PC Time-Correlated Photon Counting
DCS900PC

Best suited to

Lifetime measurements from repeated weak emission signals.

Selecting a configuration

Specify the detector pulse output, excitation reference and expected count rate. Choose a window that captures the full decay, then select the bin size; the minimum bin is not the complete instrument response.

Measurement & configuration

  • Up to 65,535 acquisition channels.
  • Steady-state acquisition support.
  • USB 3.0 data connection.

Plan the signal path

  1. 1Photon + sync pulses
  2. 2Arrival-time histogram
  3. 3Lifetime analysis

Specifications

DCS900PC specifications
Minimum time bin16 ps
ChannelsUp to 65,535
Window at minimum binApproximately 1.05 µs
Maximum bin width33.5 µs
Maximum acquisition windowApproximately 2.2 s
Instantaneous saturation count rate100 Mcps
Dead time<10 ns
InterfaceUSB 3.0

Configuration note: The 100 Mcps figure is the published instantaneous saturation rate, not a recommended continuous lifetime-measurement count rate.