Back to Home

Programmable LED solar simulation

SS-L140 / SS-L300 Steady-State LED Solar Simulators

A 21-wavelength LED array covers 350–1100 nm with independently controlled channels. SS-L140 and SS-L300 provide 140 mm and 300 mm square illuminated areas for steady-state photovoltaic and materials testing.

SS-L140 / SS-L300 Steady-State LED Solar Simulators
Best for: Perovskite cells, photovoltaic devices and experiments requiring adjustable illumination spectra.
350–1100 nm
Spectral range
21
LED wavelengths
100–1200 W/m²
Irradiance
140 / 300 mm square
Illuminated area

Overview

About the SS-L140 / SS-L300

Independently driven LED channels allow the illumination spectrum and total irradiance to be adjusted in software. Saved settings, lamp-board temperature monitoring and power correction support repeated measurements.

Effective illuminated areas are specified at a 3 cm working distance. Available spectral shapes depend on the maximum output of each channel and the total permitted electrical power.

Features

  • Programmable spectral profiles and saved illumination settings
  • Constant-current drive with temperature monitoring
  • Software operation and external control buttons

Applications

  • Perovskite and tandem-cell research
  • Steady-state photovoltaic I–V measurements
  • Semiconductor and photoelectric materials testing

Technical data

Specifications & Configuration

Illumination
LED array21 wavelength channels
Spectral range350–1100 nm
Irradiance adjustment100–1200 W/m², continuous
Effective field at 3 cm140 × 140 mm (SS-L140); 300 × 300 mm (SS-L300)
Spatial non-uniformity≤ 2%
Intensity stability≤ 0.5% over 4 hours; source stability measurements follow 20 minutes of warm-up
Source lifetime≥ 10,000 hours
Dimensions
SS-L140 main enclosure270 × 270 × 575.5 mm
SS-L300 main enclosure410 × 410 × 635.5 mm

Available configurations

SS-L140
140 × 140 mm effective illuminated area at 3 cm working distance.
SS-L300
300 × 300 mm effective illuminated area at 3 cm working distance.

Download the datasheet

Measurements, specifications and configuration options.

SS-L140 / SS-L300 datasheet (PDF)

Configure the SS-L140 / SS-L300 for your experiment

Tell us your sample size, wavelength range and measurement requirements for a configuration and quotation.

Request a Quote