Back to Home
Benchtop FTIR Spectrometer · Research-grade FTIR

FI-RXF200

FI-RXF200 Research FTIR Spectrometer

A research FTIR spectrometer with resolution better than 0.25 cm⁻¹ and a range that extends from the standard 8000–350 cm⁻¹ into the near-infrared to 12800 cm⁻¹. Sources and detectors switch automatically; internal and external accessories, from TG-IR to IR microscopy, are supported.

0.25 cm⁻¹

Resolution, better than

12800–350 cm⁻¹

Range with NIR option

≥ 55,000:1

Signal-to-noise ratio

≤ 0.1 %T

Transmittance accuracy

FI-RXF200 Research FTIR Spectrometer
Best for: Research laboratories needing high resolution, mid- and near-IR on one instrument, and coupled techniques such as TG-IR, GC-IR and reaction monitoring.

Overview

About the FI-RXF200

A research FTIR spectrometer built for the varied needs of scientific users. Resolution is better than 0.25 cm⁻¹ and the spectral range extends to 12800–350 cm⁻¹; sources and detectors can be switched automatically, and both internal and external accessories are supported.

Features

Spectral resolution

  • Better than 0.25 cm⁻¹ over the full spectrum, with very high signal-to-noise ratio
  • Upgradeable to higher resolution; covers the great majority of research needs

Extended spectral range

  • Automatic switching of optical components extends the standard mid-infrared range (8000–350 cm⁻¹) into the near-infrared (12800–4000 cm⁻¹), for mid- and near-IR work on one instrument
  • In FT-NIR, samples usually need no dilution, and because glass is transparent in the near-infrared, samples are measured directly in glass vials

Detectors

  • Room-temperature DLaTGS as standard
  • High-sensitivity options: liquid-nitrogen-cooled MCT, TE-cooled MCT, liquid-nitrogen-cooled InSb and TE-cooled InGaAs, for high sensitivity and throughput
  • Every detector uses the latest dual-channel analogue-to-digital converter with 32-bit dynamic range, for lower electronic noise

Software

  • Designed for research laboratories and simple to operate
  • All operation and evaluation functions can be applied to batches of files
  • Integrated fully automatic acquisition and online real-time reaction-monitoring software, for studying intermediate steps in chemical and pharmaceutical synthesis, catalytic and electrochemical reactions

Applications

Pharmaceuticals and life sciences

  • Protein conformation and concentration
  • High-sensitivity quantification of drug substances and excipients in aqueous solution
  • Identification of microorganisms
  • Volatility and stability of pharmaceutical products, coupled with thermal analysis

Polymers and chemicals

  • Detection and characterisation of volatilisation and decomposition, coupled with thermal analysis
  • Monitoring laboratory reactions, with a mid-infrared ATR fibre probe

Surface analysis

  • Ultra-thin films and monolayers
  • Corrosion processes

Materials science

  • Emissivity of building materials
  • Characterisation of optical materials such as infrared windows and mirrors
  • Photoacoustic spectroscopy of black materials

Semiconductors and silicon

  • Thickness of non-metallic films
  • Carbon and oxygen impurities in quality control

Soil analysis

  • Nutrients and organic matter in soil
  • Soil properties

External accessories and sample-compartment options

  • Thermal-analysis–infrared (TG-IR) coupling module
  • Gas-chromatography–infrared (GC-IR) coupling accessory
  • External sample compartment, evacuated or purged with nitrogen
  • Vacuum-sealed external chamber for connection to UHV
  • Gas cells from 2 cm to 20 m
  • Integrating sphere
  • In-situ transmission accessory
  • In-situ diffuse reflectance accessory
  • Infrared microscopy accessory
  • ATR and specular reflectance accessories

Technical Specifications

FI-RXF200 Specifications

Performance

Spectral resolution≤ 0.25 cm⁻¹
Spectral range8000–350 cm⁻¹ standard; 12800–4000 cm⁻¹ optional
Wavenumber accuracy0.005 cm⁻¹
Signal-to-noise ratio≥ 55,000:1
Transmittance accuracy≤ 0.1 %T
InterferometerCube-corner interferometer

Standard and optional configuration

ItemStandardOptional
Spectral resolution≤ 0.25 cm⁻¹—
Spectral range8000–350 cm⁻¹12800–4000 cm⁻¹
SourceCeramic sourceTungsten lamp
BeamsplitterGe-coated KBrQuartz, CaF₂, ZnSe and others
DetectorRoom-temperature DLaTGSLiquid-nitrogen-cooled MCT; TE-cooled MCT; liquid-nitrogen-cooled InSb; liquid-nitrogen-cooled Ge; TE-cooled InGaAs; and others
Sample compartmentTransmission sample compartmentInfrared emission port; external sample compartment; vacuum-sealed external chamber for UHV; and others

Datasheet

Full specification sheet for the FI-RXF200.

Configure the FI-RXF200 for your samples

Tell us what you need to measure and we will put together the configuration and a quotation.