FI-RXF200
FI-RXF200 Research FTIR Spectrometer
A research FTIR spectrometer with resolution better than 0.25 cm⁻¹ and a range that extends from the standard 8000–350 cm⁻¹ into the near-infrared to 12800 cm⁻¹. Sources and detectors switch automatically; internal and external accessories, from TG-IR to IR microscopy, are supported.
0.25 cm⁻¹
Resolution, better than
12800–350 cm⁻¹
Range with NIR option
≥ 55,000:1
Signal-to-noise ratio
≤ 0.1 %T
Transmittance accuracy

Overview
About the FI-RXF200
A research FTIR spectrometer built for the varied needs of scientific users. Resolution is better than 0.25 cm⁻¹ and the spectral range extends to 12800–350 cm⁻¹; sources and detectors can be switched automatically, and both internal and external accessories are supported.
Features
Spectral resolution
- Better than 0.25 cm⁻¹ over the full spectrum, with very high signal-to-noise ratio
- Upgradeable to higher resolution; covers the great majority of research needs
Extended spectral range
- Automatic switching of optical components extends the standard mid-infrared range (8000–350 cm⁻¹) into the near-infrared (12800–4000 cm⁻¹), for mid- and near-IR work on one instrument
- In FT-NIR, samples usually need no dilution, and because glass is transparent in the near-infrared, samples are measured directly in glass vials
Detectors
- Room-temperature DLaTGS as standard
- High-sensitivity options: liquid-nitrogen-cooled MCT, TE-cooled MCT, liquid-nitrogen-cooled InSb and TE-cooled InGaAs, for high sensitivity and throughput
- Every detector uses the latest dual-channel analogue-to-digital converter with 32-bit dynamic range, for lower electronic noise
Software
- Designed for research laboratories and simple to operate
- All operation and evaluation functions can be applied to batches of files
- Integrated fully automatic acquisition and online real-time reaction-monitoring software, for studying intermediate steps in chemical and pharmaceutical synthesis, catalytic and electrochemical reactions
Applications
Pharmaceuticals and life sciences
- Protein conformation and concentration
- High-sensitivity quantification of drug substances and excipients in aqueous solution
- Identification of microorganisms
- Volatility and stability of pharmaceutical products, coupled with thermal analysis
Polymers and chemicals
- Detection and characterisation of volatilisation and decomposition, coupled with thermal analysis
- Monitoring laboratory reactions, with a mid-infrared ATR fibre probe
Surface analysis
- Ultra-thin films and monolayers
- Corrosion processes
Materials science
- Emissivity of building materials
- Characterisation of optical materials such as infrared windows and mirrors
- Photoacoustic spectroscopy of black materials
Semiconductors and silicon
- Thickness of non-metallic films
- Carbon and oxygen impurities in quality control
Soil analysis
- Nutrients and organic matter in soil
- Soil properties
External accessories and sample-compartment options
- Thermal-analysis–infrared (TG-IR) coupling module
- Gas-chromatography–infrared (GC-IR) coupling accessory
- External sample compartment, evacuated or purged with nitrogen
- Vacuum-sealed external chamber for connection to UHV
- Gas cells from 2 cm to 20 m
- Integrating sphere
- In-situ transmission accessory
- In-situ diffuse reflectance accessory
- Infrared microscopy accessory
- ATR and specular reflectance accessories
Technical Specifications
FI-RXF200 Specifications
Performance
| Spectral resolution | ≤ 0.25 cm⁻¹ |
|---|---|
| Spectral range | 8000–350 cm⁻¹ standard; 12800–4000 cm⁻¹ optional |
| Wavenumber accuracy | 0.005 cm⁻¹ |
| Signal-to-noise ratio | ≥ 55,000:1 |
| Transmittance accuracy | ≤ 0.1 %T |
| Interferometer | Cube-corner interferometer |
Standard and optional configuration
| Item | Standard | Optional |
|---|---|---|
| Spectral resolution | ≤ 0.25 cm⁻¹ | — |
| Spectral range | 8000–350 cm⁻¹ | 12800–4000 cm⁻¹ |
| Source | Ceramic source | Tungsten lamp |
| Beamsplitter | Ge-coated KBr | Quartz, CaF₂, ZnSe and others |
| Detector | Room-temperature DLaTGS | Liquid-nitrogen-cooled MCT; TE-cooled MCT; liquid-nitrogen-cooled InSb; liquid-nitrogen-cooled Ge; TE-cooled InGaAs; and others |
| Sample compartment | Transmission sample compartment | Infrared emission port; external sample compartment; vacuum-sealed external chamber for UHV; and others |
Datasheet
Full specification sheet for the FI-RXF200.
Compare