Back to Home
Omni-λ3018i

Representative image of the product family.

View full-size image

Monochromators and SPE · Imaging Monochromator

Omni-λ3018i

Imaging Monochromator | IR coating based on Omni-λ3008i

View quote list

Product range

Where Omni-λ3018i fits

Omni-λ3018i is listed in the Imaging Monochromator family. The local catalog record provides these first selection fields: Focal length: 320 mm; Aperture ratio: f/4.2; Resolution (PMT): 0.08 nm.

Specifications

Focal length320 mm
Aperture ratiof/4.2
Resolution (PMT)0.08 nm
Resolution (CCD, 26 µm)0.173 nm
Dispersion2.29 nm/mm
Measurement conditions1200 g/mm grating at 435.8 nm, 10 µm slit width; PMT at 4 mm slit height, CCD with 26 µm pixels
Wavelength accuracy±0.2 nm
Wavelength repeatability±0.025 nm
Step size0.005 nm
Stray light1×10⁻⁵
Focal plane size30 (w) × 14 (h) mm
Optical axis height146 mm
SlitWidth 0.01–3 mm, motorised or manual; height 2, 4 or 14 mm selectable
Grating size68 × 68 mm
Grating mountInterchangeable triple grating turret
InterfaceUSB 2.0 and RS232
Dimensions420 × 295 × 232 mm
Weight16 kg

Product comparison

Check a reference product

Considering Omni-λ3018i for an existing setup? Send us the reference product number and your required specifications. Our engineers can help check the function, configuration, dimensions and interfaces.

Ask our engineers →

System integration

Integration and compatibility

Use these checks to connect Omni-λ3018i to the surrounding optical, mechanical and control system. Every value shown is taken from this model's own specification record.

Spectral range and resolution
Resolution (PMT): 0.08 nm · Resolution (CCD, 26 µm): 0.173 nm
Resolution is a budget, not a property: focal length, groove density, slit width, diffraction order and detector pixel pitch all enter it, and the quoted figure applies only at the wavelength and slit width it was measured at.
Grating and dispersion
Focal length: 320 mm · Dispersion: 2.29 nm/mm
Confirm the installed gratings, their blaze wavelength and the turret population. A shared chassis name does not imply the same grating set, and blaze position decides efficiency far more than groove density alone.
Slits and optical ports
Aperture ratio: f/4.2 · Slit: Width 0.01–3 mm, motorised or manual; height 2, 4 or 14 mm selectable
Match the input f-number to the instrument, not the other way round. Overfilling the acceptance cone adds stray light and no signal; underfilling wastes throughput you have already paid for in aperture.
Detector interface
Focal plane size: 30 (w) × 14 (h) mm · Interface: USB 2.0 and RS232
Match the detector active height and pixel pitch to the focal-plane dispersion before committing to a mount, then confirm wavelength calibration, scan repeatability and trigger behaviour for the acquisition you intend.

Technical FAQ

Integration questions for Omni-λ3018i

What spectral range and resolution does the Omni-λ3018i cover?

Resolution (PMT): 0.08 nm · Resolution (CCD, 26 µm): 0.173 nm. Resolution is a budget, not a property: focal length, groove density, slit width, diffraction order and detector pixel pitch all enter it, and the quoted figure applies only at the wavelength and slit width it was measured at.

Which gratings does the Omni-λ3018i take?

Focal length: 320 mm · Dispersion: 2.29 nm/mm. Confirm the installed gratings, their blaze wavelength and the turret population. A shared chassis name does not imply the same grating set, and blaze position decides efficiency far more than groove density alone.

What slit and port configuration does the Omni-λ3018i have?

Aperture ratio: f/4.2 · Slit: Width 0.01–3 mm, motorised or manual; height 2, 4 or 14 mm selectable. Match the input f-number to the instrument, not the other way round. Overfilling the acceptance cone adds stray light and no signal; underfilling wastes throughput you have already paid for in aperture.

Which detectors mount to the Omni-λ3018i?

Focal plane size: 30 (w) × 14 (h) mm · Interface: USB 2.0 and RS232. Match the detector active height and pixel pitch to the focal-plane dispersion before committing to a mount, then confirm wavelength calibration, scan repeatability and trigger behaviour for the acquisition you intend.

Model selection

Nearby series comparison

Compare Omni-λ3018i with the closest available models in the Imaging Monochromator family.

Selection parameterOmni-λ3018i (current)Omni-λ3017iOmni-λ3024i
Spectral range±0.2 nm±0.2 nm±0.2 nm
Resolution0.08 nm0.08 nm0.08 nm
Focal length / f-number320 mm320 mm320 mm
Grating68 × 68 mm68 × 68 mm68 × 68 mm