Opto-Mechanics · Cryogenic Piezo Stages - Thermal links
FTC16.L**
Flexible thermal connection set for 16 mm Linear and Scanner series stacks; 30 mK to 420 K, ambient/HV/UHV compatible, usable up to 35 T; choose 35, 65, or 100 mm copper coupling foil length.
Datasheet (PDF) ↗Specifications
Full specifications for this product are available on request. Contact us for the datasheet.
Toepassingen
Koelen van een bewegend monster
Voer warmte af van een tafel die moet transleren, zonder de slag te beperken.
Beheersen van piëzo-zelfopwarming
Voer de dissipatie van aangedreven assen af zodat de basistemperatuur tijdens scannen behouden blijft.
Spanningsvrije thermische verankering
Flexibele vlecht ontkoppelt thermisch contact van mechanische belasting op de positioneerderstapel.
Specifying this part
What this model has to survive, and what it needs around it
Heat sinking a stage that has to move
High-purity copper braid ties the moving body to the cold plate, conducting heat away from the stack without restraining travel or introducing mechanical strain into the positioner.
Heat load at the cold plate
Stick-slip inertial drives dissipate energy per step and draw no holding current once stopped, so the steady-state load on the mixing chamber is the wiring loom rather than the motor. Anchor the loom at each temperature stage and thermalise the moving body with an FTC copper braid.
Non-magnetic construction
Titanium bodies and BeCu fasteners throughout, rated for operation to 35 T and down to 30 mK with the .ULT option, at 2 × 10⁻¹¹ mbar with .UHV. A single steel screw substituted at installation is enough to compromise a high-field assembly.
Full series comparison, controller pairing and mounting hardware on the cryogenic piezo stages overview, or work through the selection with the configurator.
Related products
FTC25.L**
Flexible thermal connection set for 25 mm Linear and Scanner series stacks; 30 mK to 420 K, ambient/HV/UHV compatible, usable up to 35 T; choose 35, 65, or 100 mm copper coupling foil length.
FTC35.L**
Flexible thermal connection set for 35 mm Linear, Goniometer, and Scanner series stacks; 30 mK to 420 K, ambient/HV/UHV compatible, usable up to 35 T; choose 35, 65, or 100 mm copper coupling foil length.