Technology guide
Optical tweezers: trapping, detection and calibration
Separate the trapping field, position readout and calibration needed to measure microscopic forces or mechanical response.
- 01Focused laser
- 02Trapped probe
- 03Motion detection
- 04Calibration
How it works
- A tightly focused optical field can exert forces on a microscopic object. Suitable optical and particle conditions can produce a restoring trap around a stable position.
- The trapping beam and position-detection arrangement are separate system functions. A quadrant-detector board alone does not establish a complete optical trap.
- Probe size, material, medium and optical geometry influence trapping and readout. Laser output power by itself does not define trap stiffness.
- A calibrated relationship is needed to turn recorded displacement into force or material response. Detector bandwidth and digital acquisition must support the relevant motion.
What to compare
| Parameter | Check in the specification |
|---|---|
| Trapping optics | Match wavelength, objective and available optical power at the sample. |
| Probe and medium | Specify particle properties and the sample environment. |
| Detection | Check position sensitivity, optical alignment and analogue bandwidth. |
| Calibration | Define the method and physical conditions used for displacement and trap stiffness. |
Integration
- Confirm which laser, microscope, detector and acquisition electronics are included in a supplied system.
- Keep any working-prototype status and documented operating limits explicit.
- Use experimental calibration rather than inferring force sensitivity from a laser or ADC specification alone.
Related products
References
Discuss the configuration. Send the sample format, observation method and optical-interface requirements.