Optical tweezers: trapping, detection and calibration

Separate the trapping field, position readout and calibration needed to measure microscopic forces or mechanical response.

  1. 01Focused laser
  2. 02Trapped probe
  3. 03Motion detection
  4. 04Calibration
Conceptual operating path.

How it works

  • A tightly focused optical field can exert forces on a microscopic object. Suitable optical and particle conditions can produce a restoring trap around a stable position.
  • The trapping beam and position-detection arrangement are separate system functions. A quadrant-detector board alone does not establish a complete optical trap.
  • Probe size, material, medium and optical geometry influence trapping and readout. Laser output power by itself does not define trap stiffness.
  • A calibrated relationship is needed to turn recorded displacement into force or material response. Detector bandwidth and digital acquisition must support the relevant motion.

What to compare

ParameterCheck in the specification
Trapping opticsMatch wavelength, objective and available optical power at the sample.
Probe and mediumSpecify particle properties and the sample environment.
DetectionCheck position sensitivity, optical alignment and analogue bandwidth.
CalibrationDefine the method and physical conditions used for displacement and trap stiffness.

Integration

  • Confirm which laser, microscope, detector and acquisition electronics are included in a supplied system.
  • Keep any working-prototype status and documented operating limits explicit.
  • Use experimental calibration rather than inferring force sensitivity from a laser or ADC specification alone.