Back to Home
TBG45-60-ST542 Motorized Goniometer

TBG45-60 base stage shown; the requested motor or cover option is not illustrated.

View full-size image

Motorized Stages

TBG45-60-ST542 Motorized Goniometer

  • Repositioning accuracy: ≤±0.005°
  • Horizontal load capacity: 4 kg
  • Table size: 60×60 mm
  • Rotation centre height: 45 mm
View quote list
TBG Series Motorized Goniometers

Specifications

Mechanical

Rotation centre height (mm)45
Table size (mm)60×60
Angle range (°)±10
Worm gear ratio159:1
Guide mechanismArc V shaped ball guide
Worm gear materialWear-resistant tin bronze
Worm material and treatmentSteel, high-frequency quench
Body material and finishBlack anodized aluminum alloy
Thickness (mm)35

Accuracy

Repositioning accuracy (°)≤±0.005
Static clearance (µm)≤7
Backlash (°)≤0.01
Swing accuracy at rotation centre (mm)≤0.01

Sensors & connection

Limit sensors (built-in)2×GP1S092HCPI (SHARP, Japan)
Origin sensor (built-in)1×GP1S092HCPI (SHARP, Japan)
Sensor supply voltageDC5~24V±10%
Sensor outputNPN open-collector output, DC5~24V 8mA or less Residual voltage 0.3V or less (when load current is 2mA)
Sensor output logicOutput ON when sensor is blocked

Load capacity

Horizontal load capacity (kg)4
Vertical load capacity (kg)1.5
Inverted load capacity (kg)2

Construction details

  • Guidance assembly: Arc V shaped ball guide.
Product datasheet (PDF)

Downloads

Product comparison

Compare with Thorlabs GNM2/M

TBG45-60-ST542 and Thorlabs GNM2/M offer options for motorized virtual-axis tilt and goniometric alignment. Compare the listed dimensions and interfaces for your setup.

Reference configuration: ±50° virtual-axis goniometer with microstepping motor, optical encoder and integrated controller.

Published specifications for TBG45-60-ST542 and Thorlabs GNM2/M
FeatureAvailable from PrecisometerTBG45-60-ST542Comparison referenceThorlabs GNM2/M
Angular travel (°)±10±50
Moving platform (mm)60×6086.0 × 94.0
Motor and controllerFive-phase 42 mm stepper; TMC-USB-1-S528 matching controller sold separately; configure for five-phase 42 stepping motor; verify required encoder support and ordered accessoriesMicrostepping motor, optical encoder and integrated USB/Kinesis controller
Feedback configurationOutput-position encoder feedback not specified for this ordering codeIntegrated optical encoder; closed-loop operation
Pivot height above moving platform (mm)4570.00 ± 0.01
Published centered/platform load (kg)46.10 (60 N)

Similar application. Check the interfaces for your setup.

Both provide single-axis motorized rotation about a virtual point above the moving platform. Pivot height, angular range, platform, load geometry, drive and feedback differ. Stackability and a shared pivot between different brands are not established; check both drawings for the intended assembly. The five-phase option is reviewed separately and does not inherit base-model speed or step-size specifications. Precisometer requires its matching separately ordered controller and drive configuration; GNM2/M includes a controller, power supply, USB cable and trigger cables. Controller wiring, feedback, stack interfaces and mounting are not cross-brand interchangeable by assumption. This comparison covers the listed features and does not establish interchangeability or identical performance.

Comparison references: TBG45-60-ST542 catalogue specifications and Thorlabs GNM2/M published specifications. Checked 20 October 2026. Thorlabs is a trademark of its respective owner. Precisometer is not affiliated with or endorsed by Thorlabs.

All Thorlabs part-number comparisons →

Controllers for TBG45-60-ST542

For the five-phase 42 stepping motor this stage uses.