Spectroscopy Detectors · Silicon photodetectors
DSi200 / DSi300
DSi Silicon Photodetectors
- Models: DSi200 (UV enhanced) / DSi300 (blue enhanced)
- Effective area: 100 mm² (Ø11.28 mm)
- Spectral range: DSi200: 200–1100 nm; DSi300: 350–1100 nm
- Peak wavelength: 820 nm / 970 nm
- Peak responsivity: 0.52 A/W / 0.60 A/W
- Operating temperature: −10 to +60 °C
Cataloguscontext
De plaats van DSi200 / DSi300 in het assortiment
DSi200 / DSi300 staat als DSi Silicon Photodetectors vermeld binnen de familie Silicon photodetectors. Het lokale catalogusrecord bevat als eerste selectievelden: Models: DSi200 (UV enhanced) / DSi300 (blue enhanced); Effective area: 100 mm² (Ø11.28 mm); Spectral range: DSi200: 200–1100 nm; DSi300: 350–1100 nm.
Specifications
| Models | DSi200 (UV enhanced) / DSi300 (blue enhanced) |
|---|---|
| Effective area | 100 mm² (Ø11.28 mm) |
| Spectral range | DSi200: 200–1100 nm; DSi300: 350–1100 nm |
| Peak wavelength | 820 nm / 970 nm |
| Peak responsivity | 0.52 A/W / 0.60 A/W |
| Operating temperature | −10 to +60 °C |
| Signal output | Current |
Volledige bronspecificatie
Vergelijking per model
De volledige vergelijking van de fabrikant staat hieronder, zodat modelverschillen zichtbaar blijven en niet tot één bereik worden teruggebracht.
| Parameter | DSi200 — UV enhanced | DSi300 — blue enhanced |
|---|---|---|
| Effective area | 100 mm² (Ø11.28 mm) | 100 mm² (Ø11.28 mm) |
| Spectral range | 200–1100 nm | 350–1100 nm |
| Peak wavelength | 820 nm | 970 nm |
| Peak responsivity | 0.52 A/W | 0.60 A/W (> 0.55) |
| Specified responsivity | 0.14 A/W (> 0.09) at 254 nm | 0.20 A/W (> 0.15) at 410 nm |
| Rise time | 5.9 µs | 2 µs |
| Operating temperature | −10 to +60 °C | −10 to +60 °C |
| Storage temperature | −20 to +70 °C | −20 to +70 °C |
| Shunt resistance | 10 MΩ (> 5 MΩ) | > 10 MΩ |
| Noise-equivalent power | 4.5 × 10⁻¹³ W/√Hz | 2.0 × 10⁻¹³ W/√Hz |
| Maximum operating current at 0 V bias | 0.1 mA | 10.0 mA |
| Junction capacitance at 0 V bias | 4500 pF | 8800 pF |
| Signal output | Current | Current |
| Output polarity | P | P |
Ontwerp- en integratiedetails
- Large 100 mm² active area in UV-enhanced and blue-enhanced versions.
- BNC current output and direct mounting flange for Omni-λ monochromator slit assemblies.
- ZAMP transimpedance amplifier recommended when voltage output or improved sensitivity is required.
- Direct current-input connection is possible when the downstream instrument supports it.
Toepassings- en aansluitnotities
- Use a ZAMP preamplifier with oscilloscopes, lock-in amplifiers or DCS103 acquisition when current-to-voltage conversion is needed.
- DCS300PA already integrates a signal amplifier, so an additional preamplifier is not required.
Waarden zijn overgenomen uit de gearchiveerde producttabel en tekeningen. Bevestig de uiteindelijke detector, elektronica en spectrograafaansluiting voor uw configuratie.