Power, wavelength, beam profile, safety
HP50 Laser Power Meter
- Model: HP50
- Power Range: 0.5 W - 50 W
- Active Area Diameter: 25 mm
- Wavelength Range: 0.19-25 μm
- Max. Power Density: 40 kW/cm²
- Max. Energy Density: 0.15 J/cm²,1064 nm,1 ns

Model Specifications
HP50 Laser Power Meter
Model specifications
| Parameter | Specification |
|---|---|
| Model | HP50 |
| Power Range | 0.5 W - 50 W |
| Active Area Diameter | 25 mm |
| Wavelength Range | 0.19-25 μm |
| Max. Power Density | 40 kW/cm² |
| Max. Energy Density | 0.15 J/cm²,1064 nm,1 ns |
| Measurement Uncertainty | ±2% |
| Linearity | ±1% |
| Resolution | 100 mW |
| Sensitivity | 500 mW |
| Accuracy | <±2% |
| Cooling Method | Air cooling |
| Temperature (Operating) | 5°C-45°C(41°F-113°F) |
| Temperature (Storage) | -20°C-70°C(-4°F-158°F) |
| Warranty Period | 1 year |
Measurement conditions
| Condition | Detail |
|---|---|
| Note 1 | Keep total optical power and average power density within the stated limits. Check the beam spot size before measurement. |
| Note 2 | The pulse-energy-density limit applies only under the listed wavelength, pulse-duration and repetition-rate conditions. It is separate from the average-power-density limit. |
Sensor, display and mechanical views

Need this configured?
Share measurement range, detector needs, interface requirements, mounting constraints and calibration needs, and we will qualify the matching configuration.