Back to Home

Power, wavelength, beam profile, safety

HP2 Laser Power Meter

  • Model: HP2
  • Power Range: 10 mW - 2 W
  • Active Area Diameter: 25 mm
  • Wavelength Range: 0.19-25 μm
  • Max. Power Density: 15 kW/cm²
  • Max. Energy Density: 0.15 J/cm²,1064 nm,1 ns
HP2 Laser Power Meter

Model Specifications

HP2 Laser Power Meter

Parent product page

Model specifications

ParameterSpecification
ModelHP2
Power Range10 mW - 2 W
Active Area Diameter25 mm
Wavelength Range0.19-25 μm
Max. Power Density15 kW/cm²
Max. Energy Density0.15 J/cm²,1064 nm,1 ns
Measurement Uncertainty±2%
Linearity±1%
Resolution100 μW
Sensitivity10 mW
Accuracy<±2%
Cooling MethodAir cooling
Temperature (Operating)5°C-45°C(41°F-113°F)
Temperature (Storage)-20°C-70°C(-4°F-158°F)
Warranty Period1 year

Measurement conditions

ConditionDetail
Note 1Keep total optical power and average power density within the stated limits. Check the beam spot size before measurement.
Note 2The pulse-energy-density limit applies only under the listed wavelength, pulse-duration and repetition-rate conditions. It is separate from the average-power-density limit.
Note 3The HP2 model document specifies 15 kW/cm² maximum power density. Use this model-specific limit rather than the higher value in the handheld-series overview.

Sensor, display and mechanical views

Mechanical outline and sensor dimensions (mm)
Mechanical outline and sensor dimensions (mm)

Need this configured?

Share measurement range, detector needs, interface requirements, mounting constraints and calibration needs, and we will qualify the matching configuration.

View enquiry list