Technology guide
Tungsten-halogen lamps: thermal spectra and stability
Use a heated-filament source with the correct electrical conditions and collection geometry for broadband measurements.
- 01Electrical heating
- 02Tungsten filament
- 03Thermal radiation
- 04Optical delivery
How it works
- A tungsten-halogen lamp emits thermal radiation from a heated filament. Its spectral distribution changes with filament temperature.
- The halogen-filled envelope supports operation of the lamp's halogen cycle under its intended conditions. It does not make the source spectrally flat.
- Electrical operating conditions influence brightness, spectrum and lifetime. A power setting is not an independent optical intensity control with a fixed spectral shape.
- Collection optics, windows and filters determine the useful delivered output and heat load on downstream components.
What to compare
| Parameter | Check in the specification |
|---|---|
| Spectral band | Include optics and detector response when selecting the useful measurement range. |
| Electrical conditions | Use the rated supply conditions and any specified regulation method. |
| Source geometry | Match the filament image and optical acceptance. |
| Stability | Keep warm-up and observation interval with an output-stability specification. |
Integration
- For absorption measurements, take reference and sample data under equivalent source conditions.
- Check whether attenuation should be optical when constant spectral shape matters.
- Confirm lamp housing, ventilation and replacement procedure for the actual assembly.
Related products
References
Discuss the configuration. Send the spectrum, sample area, delivered output and timing requirements.