Pneumatic microinjection: pressure, timing and delivery tips
Control fluid delivery through a pressure-driven tip while treating volume as a calibrated outcome.
- 01Pressure command
- 02Fluid path
- 03Delivery tip
- 04Sample
How it works
- A pneumatic injector uses a pressure difference to move fluid through a connected delivery path. A timed pressure command controls when flow is applied.
- Delivered volume depends on pressure, pulse duration, fluid properties and tip geometry. Pulse duration alone is not a universal volume calibration.
- Holding or back-pressure settings can influence leakage and sample interaction between injections. Their permitted use depends on the actual controller and tip arrangement.
- Manipulator positioning and visual observation are separate functions from pressure generation.
What to compare
| Parameter | Check in the specification |
|---|---|
| Pressure range | Match the regulated pressure and supply requirements to the delivery setup. |
| Timing | Check pulse duration, repetition and trigger behaviour. |
| Fluid path | Specify tubing, holder, tip and fluid properties. |
| Calibration | Verify delivery under the same geometry and operating conditions as the experiment. |
Integration
- Confirm whether the product includes a pressure source, regulator, tip holder and foot or external control.
- Calibrate delivered volume after changing tips or fluids.
- Keep experimental sample preparation and biological validation separate from the instrument's electrical timing specification.