X-ray sCMOS 16MP Cooled
Fibre-optic coupled 4096 × 4096 X-ray detection with air or water cooling and two standard active input sizes.
- Resolution
- 4096 × 4096 pixels
- Input sizes
- 36.9 × 36.9 mm / 66.5 × 66.5 mm
X-ray detectors
A cooled 4-megapixel detector platform with three optical input sizes for laboratory and beamline X-ray imaging.
From photon to signal
Two different responses are published: QE for the direct-detection version and CsI:Tl absorption for scintillator-coupled configurations. They are shown separately. The table’s 58% peak QE is specified at the scintillator emission wavelength without microlens; it is not an end-to-end X-ray DQE.
Select a figure to inspect the original axes, legend and curves.
QE versus photon energy for the direct-detection version only. This curve does not describe the scintillator-coupled 4MP_32, 4MP_68 or 4MP_95 configurations.
Scintillator absorption for the thicknesses in the legend. These are converter absorption curves, not complete-camera QE or DQE; the direct-detection sensor curve is shown separately.
The details behind the selection
Compare sensor configurations, coating options, operating limits and integration details directly below. Check the operating mode and temperature alongside each performance value.
| Characteristics | sCMOS 4MP_32 | sCMOS 4MP_68 | sCMOS 4MP_95 |
|---|---|---|---|
| Resolution | 2048 x 2048 | ||
| Input Size (mm) | 22.5 x 22.5 | 48 x 48 | 67.2 x 67.2 |
| Input Size (µm) | 11 x 11 | 23.4 x 23.4 | 32.8 x 32.8 |
| Frame Rate | 18 fps at full resolution | ||
| Full Well Capacity | 80,000 electrons | ||
| Read Out Noise | 1.8 electrons rms | ||
| Dark Current | <1 electron/pixel/second | ||
| Sensor Temperature (ºC) | Operating at -20°C with water cooling | ||
| Digitization | 16-bit | ||
| Peak QE | 58% at scintillator emission wavelength (without microlens) | ||
| Exposure | 50 microseconds up to 1 minute | ||
| Spatial Resolution | sCMOS_4MP_32 with Gadox:Tb(25lp/mm), sCMOS_4MP_68 with CsI(20lp/mm), sCMOS_4MP_95 with CsI(15lp/mm) | ||
| Detector Interface | Gigabit Ethernet / Genicam compliant | ||
| Energy Range | 1keV-55keV with Gadox:Tb/20keV-300keV with CsI | ||
| Parameter | Published specification |
|---|---|
| Resolution | 2048 x 2048 |
| Input Size (mm) | 22.5 x 22.5 |
| Input Size (μm) | 11 x 11 |
| Frame Rate | 18 fps at full resolution |
| Full Well Capacity | 80,000 electrons |
| Read Out Noise | 1.8 electrons rms |
| Dark Current | <1 electron/pixel/second |
| Sensor Temperature (ºC) | Operating at -20°C with water cooling |
| Digitization | 16-bit |
| Peak QE | 58% at scintillator emission wavelength (without microlens) |
| Exposure | 50 microseconds up to 1 minute |
| Spatial Resolution | sCMOS_4MP_32 with Gadox: Tb(25lp/mm), sCMOS_4MP_68 with CsI(20lp/mm), sCMOS_4MP_95 with CsI(151p/mm) |
| Detector Interface | Gigabit Ethernet / Genicam compliant |
| Energy Range | 1keV-55keV with Gadox: Tb/20keV-300keV with CsI |
| Parameter | Published specification |
|---|---|
| Resolution | 2048 x 2048 |
| Input Size (mm) | 48 x 48 |
| Input Size (μm) | 23.4 x 23.4 |
| Frame Rate | 18 fps at full resolution |
| Full Well Capacity | 80,000 electrons |
| Read Out Noise | 1.8 electrons rms |
| Dark Current | <1 electron/pixel/second |
| Sensor Temperature (ºC) | Operating at -20°C with water cooling |
| Digitization | 16-bit |
| Peak QE | 58% at scintillator emission wavelength (without microlens) |
| Exposure | 50 microseconds up to 1 minute |
| Spatial Resolution | sCMOS_4MP_32 with Gadox: Tb(25lp/mm), sCMOS_4MP_68 with CsI(20lp/mm), sCMOS_4MP_95 with CsI(151p/mm) |
| Detector Interface | Gigabit Ethernet / Genicam compliant |
| Energy Range | 1keV-55keV with Gadox: Tb/20keV-300keV with CSI |
| Parameter | Published specification |
|---|---|
| Resolution | 2048 x 2048 |
| Input Size (mm) | 67.2 x 67.2 |
| Input Size (μm) | 32.8 x 32.8 |
| Frame Rate | 18 fps at full resolution |
| Full Well Capacity | 80,000 electrons |
| Read Out Noise | 1.8 electrons rms |
| Dark Current | <1 electron/pixel/second |
| Sensor Temperature (ºC) | Operating at -20°C with water cooling |
| Digitization | 16-bit |
| Peak QE | 58% at scintillator emission wavelength (without microlens) |
| Exposure | 50 microseconds up to 1 minute |
| Spatial Resolution | sCMOS_4MP_32 with Gadox: Tb(25lp/mm), sCMOS_4MP_68 with CsI(20lp/mm), sCMOS_4MP_95 with CsI(151p/mm) |
| Detector Interface | Gigabit Ethernet / Genicam compliant |
| Energy Range | 1keV-55keV with Gadox: Tb/20keV-300keV with CSI |
Build your enquiry
Choose a published configuration, add the integration topics you want to discuss, and keep the exact model in your enquiry.
| Configuration | Pixels | Pitch / input | Cooling / acquisition | Read noise |
|---|---|---|---|---|
| 2048 × 2048 | 11 µm 22.5 × 22.5 mm | −20°C with water cooling 18 fps full frame | 1.8 e− rms | |
| 2048 × 2048 | 23.4 µm effective 48 × 48 mm | −20°C with water cooling 18 fps full frame | 1.8 e− rms | |
| 2048 × 2048 | 32.8 µm effective 67.2 × 67.2 mm | −20°C with water cooling 18 fps full frame | 1.8 e− rms |
Technical overview
A cooled 4-megapixel detector platform with three optical input sizes for laboratory and beamline X-ray imaging.
| Resolution | 2048 × 2048 pixels |
|---|---|
| Input sizes | 22.5, 48 or 67.2 mm square |
| Full-frame acquisition | 18 fps over Gigabit Ethernet |
| Read noise | 1.8 e− rms |
Explore the wider range
6 product ranges · 13 camera configurations

Fibre-optic coupled 4096 × 4096 X-ray detection with air or water cooling and two standard active input sizes.

Fibre-optic coupled 4096 × 4096 X-ray detection with a compact, brass-shielded housing optimised for short exposures.

Direct fibre-optic coupling protects the sensor while preserving detail in compact, high-resolution X-ray imaging.

A large-area X-ray detector built around four fibre-coupled 16MP sCMOS sensors.

A back-reflection detector with a central beam path for real-time Laue diffraction and crystal orientation.

Representative equipment; the complete configuration depends on your application.
An integrated Laue diffraction system for crystal orientation, with a compact format and manual positioning as standard.