Laue crystal orientation

Laue-LT

An integrated Laue diffraction system for crystal orientation, with a compact format and manual positioning as standard.

Laue-LT

Representative manufacturer photograph; housing and connectors depend on the selected configuration. Select the photograph to enlarge.

Orientation accuracy, down to
0.2°
Detector input
76 × 105 mm
Typical exposure
120–240 s typical
Energy range on sample
5–50 keV
Maximum sample envelope
50 × 50 × 26.5 mm

From photon to signal

Response for your configuration.

Full specifications

A system-level QE curve is not published. Laue selection depends on beam energy, detector input, crystal geometry, exposure and orientation accuracy. The model and positioning tables below preserve those system specifications.

The details behind the selection

Full technical specifications.

Compare sensor configurations, coating options, operating limits and integration details directly below. Check the operating mode and temperature alongside each performance value.

01

Laue family — complete system comparison

8 parameters
Laue family — complete system comparison
ParameterLaue-LTLaueLaue-XT
Typical exposure time120–240 s1–30 s1–30 s
Maximum sample size (L × D × H)50 × 50 × 26.5 mmHorizontal: 250 × 250 × 40 mm; vertical: 250 × 150 × 40 mm450 × 250 × 250 mm
Scanning areaHorizontal: 50 × 26.5 mm; vertical: 50 × 50 mmHorizontal: 250 × 40 mm; vertical: 250 × 150 mm250 × 250 mm
Beam size on sampleAdjustable down to 1 mm450 µm; 250 µm fine focus450 µm; 250 µm fine focus
Active detector input76 × 105 mm155 × 105 mm155 × 105 mm
Orientation accuracy, down to0.2°0.1°0.1°
Sample alignment cameraNot includedIncluded as standardIncluded as standard
Energy range5–50 keV5–29 keV5–29 keV
02

Laue / Laue-LT — positioning, mechanics & software comparison

25 parameters
Laue / Laue-LT — positioning, mechanics & software comparison
—LaueLaue-LT
Active Input Area (h) x (v) mm≈ 155 x 105≈ 76 x 105
Input Pixel Size of 61.1 µm square2570 x 17101285 x 1710
Detector Exposure Time Range1 ms to 15 min1 ms to 15 min
Pixel Binning Option✓✓
Typical Exposure Time Ranges1 s - 30 s120 s - 240 s
Pixel Binning from 1x1 to 8x8✓✓
Crystal Orientation Down to Accuracy0.1°0.2°
Solution from Random Si Orientation≈ 5 s≈ 5 s
Alignment Camera (top viewing of the sample)✓×
Laser Distance Measurement Tool✓Pointer arm
Motorised Axis Z: +/- 20 mm, X: +/- 25 mm, Y: +/- 100 mm adjustments with 20 microns accuracy and maximum 10kg load capacity✓Manual stages as standard; Maximum Y movement of +/- 10 mm
Automated Point-Scanning Routines✓×
Motorised Goniometer and Rotation Axis✓Manual as standard
Beam Size on the Sample450 µmAdjustable down to 1 mm
Sealed X-ray Tube with Tungsten Target✓✓
Average Spectrum on Sample (Bremsstrahlung)5 - 29 keV5 - 50 keV
PSEL X-ray Generator Fast automatic ramp up/down procedures software controlled up to 50 kV and 1 mA: ie. 50W power✓✓
Forced-Air Cooling✓✓
Requires Single Mains Power Connection✓✓
Standard Build Cabinet DimensionsW 689 x D 676 x H 800 mmW 649 x D 450 x H 874 mm
Custom Workbench Cabinet mount with space for monitor, keyboard and mouse (150 x 75 x 83 cm)✓✓
Reference Si Sample✓✓
X-ray Source Alignment Camera✓×
PSEL Market-Leading Laue Software✓✓

The comparison sheet describes the configurations and options in that revision. The newer family brochure adds Laue-XT and the sample envelopes above; confirm selected positioning and cabinet options.

Included across the Laue family

  • Sealed X-ray tube with tungsten target
  • Reference silicon sample
  • Cabinet workbench with space for the PC and controls
  • PSEL Laue crystal-orientation software

Build your enquiry

Start with your application.

Capture the system requirements and integration details you want to discuss with Precisometer.

Laue-LT configuration and enquiry

Tell us about your experiment so we can specify the system with you.

Your starting configuration

Laue-LT

Orientation accuracy, down to
0.2°
Detector input
76 × 105 mm
Typical exposure
120–240 s typical
Energy range on sample
5–50 keV
Maximum sample envelope
50 × 50 × 26.5 mm

Selected options are requirements for discussion. Compatibility and final performance are confirmed for your configuration.

Review enquiry

Discuss selection, integration and a quotation with Precisometer.

Technical overview

Designed around the measurement.

An integrated Laue diffraction system for crystal orientation, with a compact format and manual positioning as standard.

Single-crystal orientationCrystal cutting & preparationTurbine bladesWafer orientation mapping
Laue-LT technical overview
Orientation accuracy, down to0.2°
Detector input76 × 105 mm
Typical exposure120–240 s typical
Energy range on sample5–50 keV
Maximum sample envelope50 × 50 × 26.5 mm

Explore the wider range

Other routes to your measurement.

All Photonic Science products
Product selection

What do you need to measure?

Start with your signal, or search directly for a camera, sensor option or application.

6 product ranges · 16 camera configurations

X-ray sCMOS FDS
Photonic Science / X-ray detectors

X-ray sCMOS FDS

Direct fibre-optic coupling protects the sensor while preserving detail in compact, high-resolution X-ray imaging.

Standard configurations
FDS 1.77MP_17_S, 2.86MP_11_S and 7.01MP_17_S
Acquisition
10–37 fps, model dependent
3 configurations · specifications
X-ray sCMOS 4MP
Photonic Science / X-ray detectors

X-ray sCMOS 4MP

A cooled 4-megapixel detector platform with three optical input sizes for laboratory and beamline X-ray imaging.

Resolution
2048 × 2048 pixels
Input sizes
22.5, 48 or 67.2 mm square
3 configurations · specifications
X-ray sCMOS 64MP
Photonic Science / X-ray detectors

X-ray sCMOS 64MP

A large-area X-ray detector built around four fibre-coupled 16MP sCMOS sensors.

Architecture
Four 16MP sCMOS sensors with fibre-optic coupling
Acquisition
Over 5 fps full resolution; subarea / line modes available
1 configurations · specifications