Back to Home

Power stability, linewidth, beam quality

MDL-III-690L690 nm red CW laser

690 nm

MDL-III-690L is a 690 nm CW laser. Used for medical imaging, flow cytometry, dna sequencing. Laser. Key specifications: output / average power 1~250 mW, operating mode CW, power stability (rms, 4 hours ± 3°c) <2%, <1%, <0.5%. View specifications, datasheet and enquiry options from Precisometer.

Specifications

MDL-III-690L

17 specification rows

Output / average power1~250 mW
Beam / notesLaser
Operating modeCW
Power stability (rms, 4 hours ± 3°C)<2%, <1%, <0.5%
M2<1.5 / <1.2
Beam divergence, full angle (mrad)<1.0
Warm-up time (minutes)<5
Cooled methodConduction
Operating temperature (°C)10-35
Power supply (100-240VAC)PSU-III-FDA/PSU-III-LED/PSU-A-D/PSU-III-OEM
Expected lifetime (hours)>10000
Central wavelength (nm)690±5
Output power (mW)1-250
Transverse modeNear TEM00
Polarization ratio>50:1, (>100:1 optional) Horizontal±5 degree (Vertical optional)
Beam height from base plate (mm)24.8
Modulation optionDC-1kHz, 1kHz-10kHz, 10kHz-30kHz, 30kHz-100kHz optional; TTL and Analog optional

Application context

Why this design is used

Applications named on this model’s datasheet, expanded with the model’s optical specifications.

Medical imaging

MDL-III-690L: 690 nm emission can be matched to the absorption or excitation band used in Medical imaging. The stated stability or noise figure (<2%, <1%, <0.5%) helps keep measured intensity changes attributable to the sample rather than source drift.

Flow cytometry

MDL-III-690L: 690 nm emission can be matched to the absorption or excitation band used in Flow cytometry. The stated stability or noise figure (<2%, <1%, <0.5%) helps keep measured intensity changes attributable to the sample rather than source drift.

DNA sequencing

MDL-III-690L: 690 nm emission can be matched to the absorption or excitation band used in DNA sequencing. The stated stability or noise figure (<2%, <1%, <0.5%) helps keep measured intensity changes attributable to the sample rather than source drift.

System integration

Integration and compatibility

Use these checks to connect MDL-III-690L to the surrounding optical, mechanical and control system. Every value shown is taken from this model's own specification record.

Thermal management
Cooled method: Conduction · Operating temperature (°C): 10-35
Provide the stated cooling method and keep the laser-head base thermally coupled; temperature drift moves wavelength, output power and beam pointing.
Beam parameters at the output
M2: <1.5 / <1.2 · Beam divergence, full angle (mrad): <1.0
Size the downstream optics from the stated beam diameter and divergence rather than from the aperture of the housing, and confirm the polarisation state before specifying isolators or polarising optics.

Technical FAQ

Integration questions for MDL-III-690L

What cooling does the MDL-III-690L require?

Cooled method: Conduction · Operating temperature (°C): 10-35. Provide the stated cooling method and keep the laser-head base thermally coupled; temperature drift moves wavelength, output power and beam pointing.

What beam does the MDL-III-690L deliver?

M2: <1.5 / <1.2 · Beam divergence, full angle (mrad): <1.0. Size the downstream optics from the stated beam diameter and divergence rather than from the aperture of the housing, and confirm the polarisation state before specifying isolators or polarising optics.

Need this configured?

Share wavelength, operating mode, output power or pulse energy, linewidth or pulse width, delivery, control, and safety requirements and we will qualify the matching configuration and lead time.

Open laser selector

Model selection

Nearby series comparison

Compare MDL-III-690L with the closest available models in the CW Lasers family. Confirm option-dependent values on the final configuration before ordering.

Selection parameterMDL-III-690L (current)FC-D-690MLL-III-690L
Wavelength690 nm690 nm690 nm
Optical power / pulse energy1~250 mW1~450 mW1~250 mW
Operating mode / pulse widthCWConfirm for this modelCW
Beam quality<1.0Confirm for this modelNear TEM00