Back to Home
Measurement & Safety

FPSI 200-532

Longitudinal Mode Measurement

  • Model: FPSI 200-532
  • Wavelength range: 532-633 nm; other wavelengths customised
  • Free spectral range: 2.5 GHz
  • Total finesse: ≥200
  • Resolution: <12.5 MHz
  • Cavity length: 30 mm
Longitudinal Mode Measurement

How the measurement works

A sawtooth voltage changes the mirror separation. The photodiode records transmitted resonances as the cavity scans, revealing single- or multiple-longitudinal-mode structure.

Features

  • Confocal Fabry-Perot layout
  • Photodiode detector included
  • Mirror reflectance above 99% across the listed band
  • Adjustable high-voltage scan amplitude and duty cycle

Applications

  • Single-frequency laser assessment
  • Longitudinal-mode structure
  • Interferometry and narrow-line laser experiments

Technical specifications

Interferometer

ParameterSpecification
ModelFPSI 200-532
Wavelength range532-633 nm; other wavelengths customised
Free spectral range2.5 GHz
Total finesse≥200
Resolution<12.5 MHz
Cavity length30 mm
Mirror substrateUV fused silica

Scan controller

ParameterSpecification
WaveformSawtooth
Output voltage0-200 V
Listed scan-period settings0.05 / 0.1 / 0.5 / 1 / 5 s

Configure your measurement

Include these details with your enquiry so we can match the instrument and accessories.

  • Laser wavelength and expected linewidth/mode spacing
  • Input power, beam alignment and detector readout
  • Required scan period and high-voltage control