Back to Home

Power stability, linewidth, beam quality

UV-DP-313-A313 nm UV CW laser

313 nm

UV-DP-313-A is a 313 nm CW laser. Used for flow cytometry, dna sequencing, optical instrument, medical imaging. Laser. Key specifications: output / average power 1~500 mW, operating mode CW, power stability (rms, 4 hours±3℃) <3%, <2%, <1% / <2%, <1%, <0.5%. View specifications, datasheet and enquiry options from Precisometer.

Specifications

UV-DP-313-A

17 specification rows

Wavelength313 nm
Output / average power1~500 mW
Beam / notesLaser
Operating modeCW
Power stability (rms, 4 hours±3℃)<3%, <2%, <1% / <2%, <1%, <0.5%
Transverse modeNear TEM00
Beam divergence, full angle (mrad)<3.0
Warm-up time (minutes)<10
Cooled methodAir Cooled / WCH-580 (Water Cooled)
Operating temperature (℃)10-35
Power supply (100-240VAC)PSU-DP-A / /
Operating voltage (VDC)/ / 12V/ 25A
Expected lifetime (hours)>10000
Output power (mW)1-500
Beam diameter at the aperture (1/e2, mm)<3.0
Polarization ratio>50:1(>100:1 optional) Vertical (Horizontal optional)
Beam height from base plate (mm)50
Modulation option/

Application context

Why this design is used

Typical uses for this laser type, with the selection logic tied to the model record.

Flow cytometry

UV-DP-313-A: 313 nm emission can be matched to the absorption or excitation band used in Flow cytometry. The stated stability or noise figure (<3%, <2%, <1% / <2%, <1%, <0.5%) helps keep measured intensity changes attributable to the sample rather than source drift.

DNA sequencing

UV-DP-313-A: 313 nm emission can be matched to the absorption or excitation band used in DNA sequencing. The stated stability or noise figure (<3%, <2%, <1% / <2%, <1%, <0.5%) helps keep measured intensity changes attributable to the sample rather than source drift.

Optical instrument

UV-DP-313-A: 313 nm emission determines compatible optics, coatings and detectors for Optical instrument. The catalogued beam parameter (Near TEM00) helps predict how the source will focus and propagate through the instrument.

Medical imaging

UV-DP-313-A: 313 nm emission can be matched to the absorption or excitation band used in Medical imaging. The stated stability or noise figure (<3%, <2%, <1% / <2%, <1%, <0.5%) helps keep measured intensity changes attributable to the sample rather than source drift.

System integration

Integration and compatibility

Use these checks to connect UV-DP-313-A to the surrounding optical, mechanical and control system. Every value shown is taken from this model's own specification record.

Thermal management
Cooled method: Air Cooled / WCH-580 (Water Cooled) · Operating temperature (℃): 10-35
Provide the stated cooling method and keep the laser-head base thermally coupled; temperature drift moves wavelength, output power and beam pointing.
Beam parameters at the output
Transverse mode: Near TEM00 · Beam divergence, full angle (mrad): <3.0
Size the downstream optics from the stated beam diameter and divergence rather than from the aperture of the housing, and confirm the polarisation state before specifying isolators or polarising optics.

Technical FAQ

Integration questions for UV-DP-313-A

What cooling does the UV-DP-313-A require?

Cooled method: Air Cooled / WCH-580 (Water Cooled) · Operating temperature (℃): 10-35. Provide the stated cooling method and keep the laser-head base thermally coupled; temperature drift moves wavelength, output power and beam pointing.

What beam does the UV-DP-313-A deliver?

Transverse mode: Near TEM00 · Beam divergence, full angle (mrad): <3.0. Size the downstream optics from the stated beam diameter and divergence rather than from the aperture of the housing, and confirm the polarisation state before specifying isolators or polarising optics.

Need this configured?

Share wavelength, operating mode, output power or pulse energy, linewidth or pulse width, delivery, control, and safety requirements and we will qualify the matching configuration and lead time.

Open laser selector

Model selection

Nearby series comparison

Compare UV-DP-313-A with the closest available models in the CW Lasers family. Confirm option-dependent values on the final configuration before ordering.

Selection parameterUV-DP-313-A (current)UV-DP-313-BMSL-FN-320
Wavelength313 nm313 nm320 nm
Optical power / pulse energy1~500 mW500~4000 mW1~30 mW
Operating mode / pulse widthCWCWCW
Beam qualityNear TEM00Near TEM00TEM00