Back to Home
← Lasers

Femtosecond · technical draft

Femtosecond fibre laser systems

Amplified femtosecond systems with free-space output and optional dispersion control.

  • 780-1560 nm selected options
  • <200 fs in listed standard configurations
  • Wavelength-specific average-power entries

Select the operating point using wavelength, pulse duration and average power together. Dispersion compensation can be specified for the optical path to the sample; optional pulse picking requires a timing review.

Configuration options

Each row is a separate configuration. Limits from different rows must not be combined.
Centre wavelengthPulse durationRepetition rateAverage powerSpectral width
780 nm<200 fs10-80 MHz>1 W>3 nm
920 nm<200 fs10-80 MHz>1 W>10 nm
1030-1080 nm<200 fs10-100 MHz>2 W>10 nm
1100-1300 nmCustom; confirmCustom; confirmCustom; confirmCustom; confirm
1178 nm<200 fs30-100 MHz>1.5 W>10 nm
1560 nm<200 fs10-80 MHz>1 W>10 nm

Technical details

Output delivery
Free space
Beam propagation factor
M² <1.1, typical
Polarization extinction ratio
>20 dB, typical

Integration and delivery

Laser head
329 x 215 x 81.5 mm
Rack unit
248 x 436 x 83 mm; built-in supply; air cooled
Unit separation
2 m

Conditions and configuration notes

  • The repetition rate and pulse duration are fixed choices within the stated intervals; those intervals are not a claim of user adjustment on every unit.
  • Average power depends on wavelength, repetition rate and pulse settings. Maxima from different options must not be combined into one operating point.
  • The 1100-1300 nm option has custom pulse, repetition-rate, bandwidth and power values; no standard numerical limits are assigned.
  • Optional dispersion compensation and acousto-optic pulse picking must be specified separately.

Applications to discuss

  • Multiphoton microscopy
  • Two-photon polymerization
  • Optical metrology

Specify your configuration

Precisometer supports selection and integration. Contact us for a quotation against an agreed configuration.

  • Centre wavelength, pulse duration and repetition rate
  • Average power and pulse requirements at the sample
  • Delivery optics, dispersion and electrical or thermal interfaces
View enquiry list →

Local technical draft · configuration and final specifications require confirmation.