Specify the power and delivery path of each channel independently. The two beams can operate at the same time; a shared enclosure does not imply that the beams are combined into one output.
Configuration options
| Channel | Wavelength | Listed output |
|---|---|---|
| Blue channel | 488 ± 0.5 nm | 1000 mW |
| Green channel | 514 ± 0.5 nm | 1500 mW |
Technical details
- Power adjustment
- 40-100%
- Emission linewidth
- 0.1 nm
- Long-term wavelength stability
- ±0.1 nm
- Beam propagation factor
- M² <1.1
- Power drift over 8 hours
- 2% peak-to-peak
- Polarization
- Linear
- Transverse mode
- TEM00
Integration and delivery
- Format
- 2U, 19-inch rack platform; separate frequency-conversion head
- Outputs
- Two laser beams
- Control
- RS232 and USB graphical interface
Conditions and configuration notes
- The listed channel powers are not a guarantee of power after additional fibre coupling or isolation.
- Confirm how the two channels are delivered and whether any beam-combining optics are included.
Applications to discuss
- Two-colour microscopy
- Flow-cytometry excitation
Specify your configuration
Precisometer supports selection and integration. Contact us for a quotation against an agreed configuration.
- Power and delivery requirements for each channel
- Whether both beams need a shared axis
- Independent channel control and modulation