Cooled sCMOS 4MP
Cooled, low-noise scientific imaging with 11 µm pixels and Camera Link or GigE acquisition.
- Resolution
- 2048 × 2048 pixels
- Pixel pitch
- 11 µm
Intensified & gated
An 18 mm image intensifier fibre-coupled to a CCD readout, with nanosecond gating and a choice of photocathode and phosphor.
From photon to signal
Compare the photocathode curves before selecting the intensifier. The full model table includes MCP gain, phosphor, dynamic range and gate repetition conditions.
Select a figure to inspect the original axes, legend and curves.
Original intensifier photocathode curves. These describe the listed photocathode materials, not the end-to-end efficiency of the complete fibre-coupled camera. The chosen photocathode, MCP gain and phosphor must be specified together.
The details behind the selection
Compare sensor configurations, coating options, operating limits and integration details directly below. Check the operating mode and temperature alongside each performance value.
| Characteristics | ICCD_18 | ICMOS_18 |
|---|---|---|
| Sensor Resolution | 1940 x 1460 @ 4.54µm pitch | 1604 x 1100 @ 9µm pitch |
| Frame rate (fps)) | 25 | 90 |
| Dynamic range | Intrascene 108 | Intrascene 108 |
| Intensifier input size (mm) | 18 | |
| Intensifier resolution (lp/mm) | 60 single MCP, 30 dual MCP | |
| Photocathodes | Bi-alkali, Multi-alkali (Gen 2) and GaAs (Gen 3) | |
| Intensifier gain (single MCP) | 104 | |
| Intensifier gain (dual MCP) | 106 | |
| Phosphor screen | P43 or P46 | |
| Minimum gate speed (ns) | 3 | |
| Gate repetition rate (kHz) | 300 | |
| In camera corrections | bright pixel, offset, flatfield | |
| Camera interface | GigE,HD-SDI | |
The supplied Issue 4 specifications take precedence over the earlier web table: ICCD uses 1940 × 1460 pixels at 4.54 µm and 25 fps; ICMOS uses 1604 × 1100 at 9 µm and 90 fps. Intensifier selection determines spectral response.
Build your enquiry
Choose a published configuration, add the integration topics you want to discuss, and keep the exact model in your enquiry.
Sensor format and acquisition rate follow the supplied Issue 4 datasheet. Earlier manufacturer web tables show a different CCD configuration; specify the required sensor and timing in your enquiry.
Technical overview
An 18 mm image intensifier fibre-coupled to a CCD readout, with nanosecond gating and a choice of photocathode and phosphor.
| Intensifier input | 18 mm |
|---|---|
| Minimum gate | 3 ns |
| Gate repetition | Up to 300 kHz |
| Readout | 1940 × 1460 pixels; 25 fps |
Explore the wider range
6 product ranges · 8 camera configurations

Cooled, low-noise scientific imaging with 11 µm pixels and Camera Link or GigE acquisition.

Cooled, low-noise scientific imaging with a 36.9 mm square sensor, 9 µm pixels and interface-specific acquisition modes.

Low-noise sCMOS imaging from daylight to starlight, with a compact, low-SWaP HD format.

Low-noise sCMOS imaging from daylight to starlight, with a 9.4MP sensor and 5GigE interface.

An 18 mm image intensifier fibre-coupled to a CMOS readout, with nanosecond gating and a choice of photocathode and phosphor.

Representative equipment; the complete configuration depends on your application.
Custom camera and detector integration from sensor selection and cooling through mechanics, interfaces and software.